Immunity Test

IEC 60601

Test level

Compliance level Electromagnetic

environment –

guidance

Surge

IEC

61000-4-5

± 1 kV line(s) to

line(s)

± 2 kV line(s) to

earth

Not

applicable *)

Not

applicable *)

Not

applicable *)

Voltage dips,

short interrup-

tions and voltage

variations on

power supply in-

put lines

IEC

61000-4-11

<5 % UT

(>95 % dip

in UT) for

0.5 cycle

40 % UT

(60 % dip

in UT) for 5 cycles

70 % UT

(30 % dip

in UT) for 25 cy-

cles

<5 % UT

(>95 % dip

in UT) for 5 s

Not

applicable *)

Not

applicable *)

34

Summary of Contents for D-m CB

Page 1: ...USER INSTRUCTIONS THE WIDEX DREAM FAMILY D m CB model Behind the ear...

Page 2: ...HEARING AID To be filled out by the hearing care professional Your hearing aid series PROGRAMS Master Music TV Comfort Phone Zen SPECIAL PROGRAMS Master Zen Master Phone SMARTTOGGLE PROGRAMS Zen Phon...

Page 3: ...NOTE Read this booklet and the booklet Ear sets for Widex hearing aids carefully before your start using your hearing aid 3...

Page 4: ...use 9 The battery 9 Sound signals 11 How to tell right from left 12 Turning the hearing aid on and off 12 Putting on and removing your hearing aid 13 Sound adjustment 14 Programs 15 Using a telephone...

Page 5: ...EC directives 47 FCC and IC statements 47 SYMBOLS 51 5...

Page 6: ...E Your hearing aid and its accessories may not look exactly as illustrated in this booklet We also reserve the right to make any changes we consider necessary Your hearing aid at a glance If you need...

Page 7: ...the ear canal and prevent ear infections Contact your doctor or hearing care professional immediately if you sus pect you may have an ear infection Remove your hearing aids before showering swimming...

Page 8: ...f 132 dB There may therefore be a risk of damaging your remain ing hearing Do not expose your hearing aids to extreme temperatures or high humid ity and dry them quickly if they get wet or if you pers...

Page 9: ...ined in hearing re habilitation Intended use The hearing aids are intended as air conduction amplification devices to be used in everyday listening environments The hearing aids may be equipped with t...

Page 10: ...could leak and ruin your hearing aid Dispose of used batteries as indicated on the packaging and take note of the expiry date Low battery indication When the battery is weak a sound signal will play...

Page 11: ...tery NOTE Avoid dropping your hearing aid hold the hearing aid above a soft surface while changing the battery Sound signals Your hearing aid plays sounds to inform you that certain features have been...

Page 12: ...hat warns you whenever it loses contact with the hearing aid in the opposite ear You will hear a spoken message in your ear How to tell right from left The hearing aid for your right ear has a red mar...

Page 13: ...g your hearing aid To put on your hearing aid 1 2 1 Insert the ear set in the ear while holding the lower part of the tube Pulling the outer ear up wards and backwards at the same time can be helpful...

Page 14: ...ound environment If you have a remote control you can also adjust the sound manually de pending on your needs and preferences Depending on the hearing aid settings and features any sound adjustment yo...

Page 15: ...am if you want to listen to music TV This program is ideal when you want to listen to the TV Comfort This program reduces background noise Reverse focus Use this program if you want to focus on sounds...

Page 16: ...asily change your program selection Changing programs To change programs simply push the program key on your remote control Zen To access this program press and hold the program key for more than one...

Page 17: ...an be set by your hearing healthcare pro fessional according to your needs Benefits The Zen program may provide a relaxing listening background for some people When the Zen program is used in a tinnit...

Page 18: ...ess mode will allow you to cycle through the available Zen styles You can exit the long press mode by pressing and holding the program button for more than one second Because of the unique ways in whi...

Page 19: ...apid progressive hearing loss within previous 90 days Acute or chronic dizziness Sudden unilateral hearing loss in previous 90 days Warnings Use of the Zen program may interfere with hearing everyday...

Page 20: ...loudness tolerance of sounds speech not as clear or worsening tinnitus Using a telephone with hearing aids When you use a phone hold it against your head at an angle above your ear rather than direct...

Page 21: ...e comfortable to wear Wipe the hearing aid with a soft cloth for example the cloth you received from your hearing care profes sional If the microphone openings are still blocked contact your hearing c...

Page 22: ...he ear set manual Do not use any kind of liquid or disinfectant to clean your hearing aid Clean and inspect your hearing aid every day after use to check that it is not broken If the hearing aid break...

Page 23: ...ening to TV and audio PHONE DEX for easy landline use FM DEX for streaming audio and FM signals UNI DEX for connecting hearing aids to mobile phones CALL DEX for easy wireless connection to mobile pho...

Page 24: ...sure the battery drawer is completely closed The battery does not work Insert a new battery The hearing aid volume is not powerful enough Your ear is blocked by earwax Contact your doctor Your hearin...

Page 25: ...ing care professional to make sure the device is matched with hearing aids You hear interrupted speech on and off from the hearing aids or no speech muted from the transmitting hearing aid a The batte...

Page 26: ...C DEX TM DEX Bluetooth NOAHlink Antenna type Inductive an tenna Inductive an tenna Inductive an tenna Embedded ceramic an tenna Antenna di mensions 1 8 mm L 4 85 mm 8 mm L 20 mm 6 mm L 8 mm NA Modulat...

Page 27: ...uency 10 6 MHz 10 6 MHz 10 6 MHz 2 4 GHz Channel Single channel radio Single channel radio Single channel radio 5 logical chan nels Bandwidth 660 kHz 15 dB 660kHz 15 dB 660kHz 15 dB 1 MHz Data rate 21...

Page 28: ..._412832_QPN_E1 Benefits The use of wireless transmission allows convenient and synchronized control of hearing aid functions The DREAM wireless hearing aids share input information between the two par...

Page 29: ...speech is presented to one side and noise the other side this coordinated action could enhance the relative loudness of the speech sounds to the noise background and improve speech understanding for s...

Page 30: ...hearing aids and with their matched exter nal devices The requirements for the quality of service QoS vary among the various components and their intended user scenarios For programming these requirem...

Page 31: ...ss Security Measures Security of the wireless signals is assured through device system design that includes Individual MAC address for each unit which is checked during each transmis sion A built in p...

Page 32: ...by electronic equip ment RF emissions CISPR 11 Class B The DREAM hearing aid is suitable for use in all establishments in cluding domestic estab lishments and those di rectly connected to the public...

Page 33: ...ronment guidance Immunity Test IEC 60601 Test level Compliance level Electromagnetic environment guidance Electrostatic dis charge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors...

Page 34: ...e s 2 kV line s to earth Not applicable Not applicable Not applicable Voltage dips short interrup tions and voltage variations on power supply in put lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cyc...

Page 35: ...tic of a typical loca tion in a typical commercial or hospital environ ment NOTE UT is the a c mains voltage prior to the application of the test level Battery powered equipment Electromagnetic immuni...

Page 36: ...nt guidance Portable and mo bile RF communi cations equip ment should be used no closer to any part of the DREAMTM series hearing aid in cluding cables than the recom mended separa tion distance cal c...

Page 37: ...ted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 Vrms Recommended separation dis tance d 1 2 P Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 V m d 1 2 P 80 MHz to 800 MHz d 2 3 P 800 MHz to 2 5 G...

Page 38: ...tput power rating of the transmitter in watts W accord ing to the trans mitter manufac turer and d is the recommended separation dis tance in meters m Field strengths from fixed RF transmitters as det...

Page 39: ...may occur in the vicini ty of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electr...

Page 40: ...al operation If abnormal performance is observed additional measures may be necessa ry such as re orienting or re locating the DREAM series hearing aid b Over the frequency range 150 kHz to 80 MHz fie...

Page 41: ...smitter W Separation distance according to frequency of transmit ter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2...

Page 42: ...ower rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situati...

Page 43: ...EMC and radio spectrum matters for Short Range Devices in the frequency range 9 kHz 25 MHz IEC 60601 1 2 2007 adapted protocol EMC emission Immunity RF and ESD Medical electrical equip ment General re...

Page 44: ...ation that represents the longest length or worst case scenario This is acceptable because of the relative small size of the device compared to the wavelength of the RF used in the test Warning to hea...

Page 45: ...be exercised in se lecting and fitting a hearing aid whose maximum sound pressure level ex ceeds 132 decibels because there may be risk of impairing the remaining hearing of the hearing aid user Impor...

Page 46: ...a trial rental or purchase option pro gram Many hearing aid dispensers now offer programs that permit you to wear a hearing aid for a period of time for a nominal fee after which you may decide if yo...

Page 47: ...elevant provisions of Directive 1999 5 EC A copy of the Declaration of Conformity according to 1999 5 EC can be found at http widex com doc N26346 Information regarding disposal Do not dispose of hear...

Page 48: ...idential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the in structions may cause harmful interference to radio co...

Page 49: ...ximum or lesser gain approved for the transmitter by Industry Canada To reduce potential radio interference to other users the antenna type and its gain should be so chosen that the equivalent isotrop...

Page 50: ...cessaire l tablissement d une communication satisfaisante Le pr sent appareil est conforme aux CNR d Industrie Canada applicables aux appareils radio exempts de licence L exploitation est autoris e a...

Page 51: ...og item number Consult instructions for use The user instructions contain important cautionary information warnings precautions and must be read before using the prod uct Warning Text marked with a wa...

Page 52: ...out in Eu ropean CE marking directives RCM mark The product complies with electrical safety EMC and radio spec trum regulatory requirements for products supplied to the Aus tralian or New Zealand mark...

Page 53: ...53...

Page 54: ...54...

Page 55: ...55...

Page 56: ...WIDEX A S Nymoellevej 6 DK 3540 Lynge Denmark www widex com Manual no 9 514 0220 041 05 CIB number CIB288 0117 Issue 2017 01 9 514 0220 041 05...