Using the iP5 to Program Inteo
Selection of Inteo Instrument Style
IN-CIC Custom CIC
Fits up to 65-70 dB Hearing Loss
IN-X
Custom ITE or ITC with High
Fits up to 70-75 dB Hearing Loss
Definition Locator
IN-9
BTE with High Definition Locator
Fits up to 70-75 dB Hearing Loss
IN-9é
Open Fit BTE with High Definition Locator
Fits up to 25-30 dB Hearing Loss below
500 Hz and 70-75 dB above 1000 Hz.
IN-19
BTE with High Definition Locator
Fits up to 95-100 dB Hearing Loss
Pre-Programming Inteo
1. Perform programming in a quiet environment.
2. Connect Inteo to iP5.
a) Select correct programming pill or adaptor boot and attach to programming cable.
b) Confirm that the programming cable is connected to iP5.
c) For CIC models remove battery door and replace door with CIC adaptor.
d) For BTE models insert a new battery.
e) Attach connector(s) to hearing aid(s).
f) Confirm proper connections between all components.
3. Insert Inteo hearing aid(s) into the client’s ear(s).
4. Turn on iP5: Press [
on R
] for right or [
on L
]
for left (See A).
Note 1: Only one ear can be fit at a time. In addition adjustments cannot be made, nor data
entered, when the hourglass is active.
Note 2: An Inteo hearing aid that has been programmed and fine tuned with the iP5 program-
mer is completely compatible with Compass. An Inteo hearing aid that has been programmed
and fine tuned with Compass
may not
be completely compatible with the iP5 programmer. If
an Inteo hearing aid with program specific fine tuning, as a result of unlinking the program
from the Master Program, is connected to the iP5 programmer all programs will be linked to
the Master program and the program specific fine tuning will be recalculated. In addition, if
on Compass the Master program has been moved to a position other than the first position the
iP5 will return it to the first position. The iP5 programmer will always ask if you want to over-
write, or recalculate, the settings according to the format that the iP5 programmer can handle.
Note 3: Conversion between Inteo IN9 and IN9 é
cannot
be accomplished on the iP5 program-
mer. This conversion can
only
be made on Compass.