SAFETY LOOP and optional INTERLOCK functionality of MPV modules
The SAFETY LOOP pins provide a closed connection between LOOP0 and LOOP1 in case the properly
wired DSUB37 connector is plugged into the MPV module. It does not have any functionality inside the
module!
The Interlock features is only available on MPV 8xxxI modules. Older or other modules that do not have
the "I" (interlock) in the part number / name , will not interlock. Furthermore the interlock will work on the
bank of 4 channels on each sub-d 37 pin connector.
Using a 5-14V source, apply +V to pin 36 and 0V to pin 17 of the MPOD module sub-d 37 pin connector
(see pin layout in chapter 2.2).
In order to use the Interlock function each channel has to be enabled for this. In MUSEcontrol Software,
right mouse click on the channel and check in the “Output Configuration” window the check mark “Enable
External Inhibit” accordingly.