12
CPT6010 Operating Instructions
Digital Pressure Transducer
CPT6010
Reading Rate
50 per second
Communications
RS-232 or RS-485. LabVIEW® drivers are available.
(LabVIEW® is a trademark of National Instruments
Corporation.)
Max Transmission
4000 feet (RS-485)
Multi-drop Capacity
The maximum number of RS-485 CPT6010 transducers which
can be connected to a single host computer is 31.
Mechanical Shock
5g max
Case Size
1.74” wide x 6.3” long (4.42 x 16 cm), not including pneumatic
and electrical connectors.
Weight
Approximately 12 ounces (28.3 grams)
Pressure Media
All media compatible with 316L stainless steel.
Fittings
Pressure Port: 1/4 inch male NPT
Reference Port: 1/16 inch barb (gauge instruments only).
Power Required
9-18 VDC, 45mA @ 12 VDC
6 pin Bendix connector #PT02E-10-6P
Options
Relief valves - up to 1000 psig
Non-standard ranges