20
CPT6010 Operating Instructions
Digital Pressure Transducer
CPT6010
Convention
Description
X
Address (single character 0 through 9 or A through Z; case insensitive)
*
Wildcard address operator
?
Query operator
<sp>
Space ( )
PW
Password; insert password in place of PW (See password card on USB
drive; case insensitive)
n
One character placeholder for digit or decimal point
n...n
Variable length field
R
Ready response
<cr>
Carriage return
<lf>
Linefeed
6.4.4 Commands and Queries
i
Notice
Each command that changes a parameter is volatile until the SAVE com-
mand is issued.
Table 6.4.4 - Command Set
Pass-
word
Protect
Command
or Query
Command String Return String
Description
No
?
#X?<cr>
X<sp>nnnnnnnnnn<cr>
<lf>
Basic query to return pressure
readings.
No
A
#XA<sp>n<cr>
R
Sets address of sensor to 0-9
or A-Z.
No
B?
#XB?<cr>
X<sp>B<sp>n<cr><lf>
Returns the turndown, cal flag
No
DC?
#XDC?<cr>
X<sp>DC<sp>nnnnnn
<cr><lf>
Returns the date of calibration
where nnnnnn is mmddyy.
Yes
DC
#XDC<sp>nnnnn
n<cr>
R
Loads the date of calibration
where nnnnnn is mmddyy.
No
FL?
#XFL?<cr>
X<sp>FL<sp>nn<cr>
<lf>
Returns the filter – it is the per-
centage of old reading added
to new reading: 0-99.
Yes
FL
#XFL<sp>nn<cr>
R
Sets the filter percentage.
No
FS?
#XFS?<cr>
X<sp>FS<sp>nnnnn<cr>
<lf>
Returns the %FS accuracy.