CPT6010 Operating Instructions
21
Digital Pressure Transducer
CPT6010
No
ID?
#XID?<cr>
X<sp>ID<sp>MENSOR
<sp>DPT6000,SN<sp>nn
nnnn,V<sp>nnnn<cr><lf>
Returns transducer ID.
N/A
PW
#XPW
R
Disables password protec-tion
for the following single com-
mand.
No
R-?
#XR-?<cr>
X<sp>R-<sp>n...n
<cr><lf>
Returns minimum range value.
No
R+?
#XR+?<cr>
X<sp>R+<sp>n...n
<cr><lf>
Returns maximum range
value.
No
SAVE
#XSAVE<cr>
R
Saves all data to EEPROM for
the current turndown.
No
SC?
#XSC?<cr>
X<sp>SC<sp>n...n
<cr><lf>
Returns a span correction
multiplier.
Yes
SC
#XSC<sp>n...
n<cr>
R
Loads a span correction
multiplier.
No
SW
#XSW<sp>n<cr>
R
Switches between ‘n’=1 for pri-
mary cal or ‘n’=2 for secondary
cal.
No
T?
#XT?<cr>
X<sp>T<sp>C<cr><lf>
Returns cal type.
No
U?
#XU?<cr>
X<sp>U<sp>n<cr><lf>
Returns pressure unit code.
See Table 6.4.5.
No
ZC?
#XZC?<cr>
X<sp>ZC<sp>n...n
<cr><lf>
Returns zero correction value
in current units.
Yes
ZC
#XZC<sp>n...
n<cr>
R
Sets zero correction value in
current units.
6.4.5 Pressure Unit Code and Conversion
Table 6.4.5 - Pressure Unit Code and Conversion
Code
Unit
PSI Conversion Factor
1
psi
1
2
inHg@0°C
2.036020
3
inHg@60°F
2.041772
4
inH2O@4°C
27.68067
5
inH2O@20°C
27.72977
6
inH2O@60°F
27.70759