WIKA operating instructions process pressure transducer, model CPT6020
0020507001F 05/2019 EN
EN
7.2.6 Calibration & Communication Software
The Operation and Calibration software utility is available on the Mensor.com website from the main menu select
Service/
Calibration
then
Download Software
then
PC Software
.
The purpose of this software is to communicate with the CPT6020 transducer. Calibration of the transducer output and
configuration of various parameters can be accomplished. It has many features that allow you to read and adjust the rich feature
set including:
• Changing filter settings
• Modifying the output format of the sensor
• Changing the pressure units output
It also has many features to adjust various calibration parameters after the appropriate password is sent:
• Setting the calibration date
• Setting the calibration interval
• Setting custom pressure units
• Setting Zero and Span
• Resetting the calibration password
• Restoring the factory calibration
Maintenance and Recalibration