30
WIKA operating instructions process pressure transducer, model CPT6020
0020507001F 05/2019 EN
EN
9. Specifications
Accuracy specifications presented herein are obtained by comparison with primary standards traceable to a national metrology
institute or recognized international standard organization. These specifications are obtained in accordance with the ISO Guide to
the Expression of Uncertainty in Measurement (GUM). The calibration program at Mensor is accredited by the American Asso-
ciation of Laboratory Accreditation (A2LA) as complying with both the ISO/IEC 17025:2005 and the ANSI/NCSL Z540-1-1994
standards. If there is an exception to the requirements and recommendations of Z540 during a calibration the exception is noted on
the individual calibration certificate.
Mensor reserves the right to change specifications without notice.
9.1 Measurement Specification
Accuracy
1)
0.020% FS
Measuring ranges
Gauge pressure
0 ... 0.36 to 0 ... 1500 psig
(0 ... 25 mbar to 0 ... 100 bar)
Bi-directional pressure
2)
± 0.18 to -15 ... 1500 psig
(± 12.5 mbar to -1 ... 100 bar)
Absolute pressure
0 ... 5 psia to 0 ... 15,015 psia
(0 ... 350 mbar to 0 ... 1001 bar abs.)
Calibration interval
185 days
Pressure units
39 and 1 user defined
CPT6020 as barometric reference
Measuring range
8 ... 17 psi abs. (552 ... 1,172 mbar abs.)
Accuracy
1)
0.020% of reading
1)
It is defined by the total measurement uncertainty, with the coverage factor (k = 2) and includes the intrinsic performance of the instrument, the measurement uncertainty of the reference
instrument, long-term stability, influence of ambient conditions, drift and temperature effects over the compensated range with recommended zero point adjustment every 30 days.
2) The negative portion of a bidirectional range has the same accuracy as the equivalent positive range.
Specifications