WIKA operating instructions, models CTH6300 and CTH6500
EN
28
2079988.08 02/2018 EN/DE/IT
Example of 1-point calibration:
1. Confirm 1-point calibration
1 P
by pressing
ENTER/MENU
.
⇒
Go
appears on the display.
2. Once the measured value is stable, confirm by pressing
ENTER/MENU
.
After approx. 2 seconds
P1
appears in the first display line for measured value
1
,
dP
appears in the 2nd line for the decimal point.
3.
Use the ▲▼ arrow keys to select the desired number of decimal places:
dP. =
two decimal places
dP . =
one decimal place (decimal point moves one position to the right)
4. Press
ENTER/MENU
to confirm the desired setting.
⇒
Si_
appears on the display.
5.
Use the ▲▼ arrow keys to select the sign:
Si_
= the number to be entered is in the negative range (below 0.00 °C)
Si
┘
= the number to be entered is in the positive range
6. Commissioning, operation