WIKA operating instructions, models CTH6300 and CTH6500
EN
29
2079988.08 02/2018 EN/DE/IT
6. Press
ENTER/MENU
to confirm the desired setting.
⇒
Fd 0
appears on the display.
7.
Use the ▲▼ arrow keys to select the range.
Fd 0
= below 1,000 °C
Fd 1
= above 1,000 °C
8. Press
ENTER/MENU
to confirm the desired setting.
⇒
00.00
appears on the display (or similar).
9. Enter the temperature of the reference at this point.
10.
Change the digits using the ▲ arrow key.
11.
Use the ▼ arrow key to switch to the next digit.
12. After completing the input of the temperature, confirm by pressing
ENTER/MENU
.
13. Use
ESC
to return to the measuring mode.
The physical calibration
oP2
cannot be cancelled using the
ESC
key. If
required, the calibration can be cancelled by switching off the measuring
instrument.
Example of the measured value display after physical calibration against a
reference standard:
The small
2
in the left corner, in combination with the display segment
CAL
in the centre
of the display, indicates that the physical probe calibration
oP2
has been activated.
6. Commissioning, operation