WIKA operating instructions, models CTH6300 and CTH6500
EN
33
2079988.08 02/2018 EN/DE/IT
6.4.7 Storage management [Lo6] (not possible for CTH6300)
1. Press the
ENTER/MENU
key and select
Lo6
using the ▲▼ arrow keys (only when the
data logger version has been ordered).
2. Press the
ENTER/MENU
key once again.
⇒
OFF
is shown in the lower line of the display.
3.
Now use the ▲▼ arrow keys to start the logger mode with
ON
.
4. Confirm the selection by pressing
ENTER/MENU
.
5.
Use the ▲▼ arrow keys to choose between the automatic Auto or manual SPot
memory.
6. Confirm by pressing
ENTER/MENU
.
7. Decide between adding or creating a new log file by selecting
Add
or
nLo6
.
⇒
When selecting the automatic memory, there is a prompt to select the measuring
interval:
1S
= 1 second
1M
= 1 minute
5S
= 5 seconds
2M
= 2 minutes
10S
= 10 seconds
5M
= 5 minutes
20S
= 20 seconds
10M = 10 minutes
30S
= 30 seconds
20M = 20 minutes
8. Confirm the input by pressing
ENTER/MENU
.
9. Use
ESC
to return to the measuring mode.
Manual saving using
SPot
. Press the
ESC
key to save individual measured values
manually.
6.5 Memory request [HOLD-MAX-MIN-AVE]
After pressing the
HOLD-MAX-MIN-AVE
key for the first time, the current measured
values at the moment of actuation are “frozen” and indicated on the display as hold
values. Press this key repeatedly to request saved maximum, minimum and average
values in the lower line (small display).
MENU
Unit
Prob
Lin2
CAL
Chnl
ArEA
Lo6
6. Commissioning, operation