24
WIKA operating instructions, models CTH63I0 and CTH65I0
EN
14120692.02 01/2019 EN/DE
When calling the
oP1
by means of
[ENTER/MENU]
the
oP1
function
(calibration by code) is activated even if the menu has been exited by
pressing
[ESC]
.
Example of the measured value display after entering a probe calibration code:
The small
1
in the left corner in connection with the display segment
CAL
in the centre of
the display indicates that the calibration by code
oP1
has been activated.
Physical calibration oP2
1. Use the
[▲]
or
[▼]
arrow keys to select
oP2
.
2. Press
[ENTER/MENU]
to confirm the desired setting.
⇒
1 P
is displayed in the lower display part.
3. Use the arrow keys
[▲]
or
[▼]
to choose from 1-point
1 P
-, 2-point
2 P
- and 3-point
3 P
- calibration.
Example of 1-point calibration:
1. Confirm 1-point calibration
1 P
by pressing
[ENTER/MENU]
.
⇒
Go
appears on the display.
2. Confirm by pressing
[ENTER/MENU]
once the measured value is stable.
After approx. 2 seconds
P1
appears in the first display line for the measured value
1
,
dP.
appears in the 2nd line for the decimal point.
oP2
1P
2P
3P
6. Commissioning, operation