WIKA operating instructions, model F1201
EN
14505091.01 08/2021 EN/DE
15
7. Maintenance and cleaning / 8. Dismounting, return ...
7. Maintenance and cleaning
For contact details see chapter 1 “General information” or the back page
of the operating instructions.
7.1 Maintenance
This instrument is maintenance-free.
Repairs must only be carried out by the manufacturer.
Only use original parts (see chapter 10 “Accessories”).
7.2 Cleaning
1. Prior to cleaning, disconnect the compression force transducer from the voltage
supply and dismount it.
2. Clean the compression force transducer with a cloth.
Electrical connections must not come into contact with moisture!
CAUTION!
Damage to the instrument
Improper cleaning may lead to damage to the instrument!
▶
Do not use any aggressive cleaning agents.
▶
Do not use any hard or pointed objects for cleaning.
7.3 Recalibration
DAkkS calibration certificate, traceable and accredited in accordance with ISO/IEC
17025
We recommend that the compression force transducer is regularly recalibrated by the
manufacturer, with time intervals of approx. 24 months. The basic settings will be cor-
rected if necessary.
8. Dismounting, return and disposal
8.1 Dismounting
Release the load from the compression force transducer, disconnect from the power
and remove from the contact surface.
8.2 Return
Strictly observe the following when shipping the instrument:
All instruments delivered to WIKA must be free from any kind of hazardous substances
(acids, bases, solutions, etc.) and must therefore be cleaned before being returned.