31
6 Set up with the display and adjustment module
WIKA Operating Instructions - Process pressure transmitter IPT-1*
▶
Basic adjustment
Display
Diagnostics
Service
Info
2. Confirm the menu "
Basic adjustment
" with
[OK]
, the menu item
"
Unit
" will be displayed.
Unit
Unit of measurement
bar
▼
Temperature unit
°C
▼
3. Activate the selection with
[OK]
and select "
Units of measure-
ment
with
[->]
.
4. Activate the selection with
[OK]
and select the requested unit with
[->]
(in the example mbar).
5. Confirm with
[OK]
and move to position correction with
[->]
.
The adjustment unit is thus switched over from bar to mbar.
Proceed as follows to select the temperature unit:
5)
1. Activate the selection with
[OK]
and select "
Temperature unit
with
[->]
.
2. Activate the selection with
[OK]
and select the requested unit with
[->]
(e.g. °F).
3. Confirm with
[OK]
.
The temperature unit is hence switched over from °C to °F.
Proceed as follows:
1. Activate in the menu item "
Position correction
" the selection with
[OK]
.
Position correction
Offset
P
=
+0000 mbar
53 mbar
2. Select with
[->]
, e.g. to accept actual measured value.
Position correction
Accept current measured
value?
▶
Accept
Edit
3. Confirm with
[OK]
and move to min.(zero) adjustment with
[->]
.
Proceed as follows:
1. Edit the mbar value in the menu item "
zero
" with
[OK]
.
Carry out a position cor-
rection
Carrying out zero adjust-
ment
5)
Selection options: °C, °F.