GB
11258421.09
01/2012
GB/D
WIKA operating instructions temperature transmitter T32.xS
19
Mounting on the measuring insert
Mount the transmitter on the circular plate of the measuring insert
using two countersunk M3 screws per EN ISO 2009. Appropriate
threaded inserts have been press-fitted in the underside of the case.
Assuming the countersinking is carried out correctly, the permissible
screw length can be calculated as follows:
Check the screw length before fixing the transmitter to the measuring
insert: insert the screw into the circular plate and verify length of 4 mm!
CAUTION!
Do not exceed the maximum permissible screw length!
The transmitter will be damaged if the screws are screwed
further than 4 mm into the bottom of the transmitter.
Mounting in connection head
Insert the measuring insert with the mounted transmitter into the
protective sheath and secure into the connecting head using screws
in pressure springs.
3173801.C
l
max.
= s + 4 mm
with
l
max.
Length of screw in mm
s Thickness of circular
plate in mm
3173801.B
max.
4
Circular plate
l
s
6. Commissioning, operation
Installation by means of DIN rail adapter
If the mechanical adapter, available as an accessory, is used the
T32.1S head transmitters can also be fixed on a DIN rail.