WIKA operating instructions Ultra High Purity transducer, model WUD-2x
11611057.05 08/2021 EN/DE
EN
4. Transport, packaging and storage
4.1 Transport
CAUTION!
Damage through improper transport
With improper transport, damage to property can occur.
▶
When unloading packed goods upon delivery as well as during internal transport,
proceed carefully and observe the symbols on the packaging.
▶
With internal transport, observe the instructions in chapter 4.2 “Packaging and storage”.
Before transport
1. Place the protection cap onto the process connection.
2. Pack the transducer in accordance with clean room specifications.
3. Protect against impacts and dropping.
After transport
▶
Check the transducer for any damage that may have been caused by transport.
Obvious damage must be reported immediately.
▶
If the instrument is transported from a cold into a warm environment, the formation of condensation
may result in instrument malfunction. Before putting it back into operation, wait for the instrument
temperature and the room temperature to equalise.
4.2 Packaging and storage
Do not remove packaging until within the clean room and just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change in installation
site, sending for repair).
Permissible conditions at the place of storage:
Storage temperature: -10 ... +60 °C [14 ... 140 °F]
Humidity: 35 ... 85 % relative humidity (no condensation)