15
WIKA operating instructions Ultra High Purity transducer, model WUD-2x
11611057.05 08/2021 EN/DE
EN
5.2.1 Setting up a voltage supply
WARNING!
Incorrect voltage supply
Explosion
▶
Follow the specifications in accordance with control drawing 14097836 (see appendix).
▶
Do not exceed a power supply of DC 30 V.
Power supply U+
DC 10 ... 30 V with output signal 4 ... 20 mA / DC 0 ... 5 V
DC 14 ... 30 V with output signal DC 0 ... 10 V
Maximum current I
max
50 mA (without switching outputs)
250 mA (with switching outputs)
Additionally for operation in hazardous areas
Ci = 42 nF (+0.3 nF/m with cable outlet)
Li = 10 μH (+2 μH/m with cable outlet)
5. Commissioning, operation