background image

WH-2600 SERIES 

 

 

 

 

 

PATIENT CARE UNITS

  INSTALLATION & OPERATION MANUAL

PAGE

 11

© Rev. 

2/2017

1.800.428.4065  

TOLL FREE 

| www.willoughby-ind.com

INSTALLATION INSTRUCTIONS (CONT.)

Step 8:  Install Faucet and Connect Water Supply

 

NOTE: Please refer to faucet manufacturer’s recommended installation instructions.

Step 9:  Completing Installation

1.) 

Connect the fl ush valve outlet to the water closet.

 

NOTE: Tighten all fl ush valve connections.

2.) 

Connect the fl ush valve water supply pipe to the fl ush valve 1” NPT inlet.

3.)  

Install Push Button Actuator plastic tubing.

4.)  

Install the P-trap (supplied by installer) to the sink drain, and then connect the P-trap to the  

 

drain pipe (piping supplied by installer).

5.)  

On models with a bedpan washer sprayer, connect sprayer hose to vacuum breaker outlet  

 

inside cabinet.

6.) 

Check all water and waste connections for leaks.

Summary of Contents for WH-2600 Series

Page 1: ...the Anchors and Backing 8 Step 3 Install the Vacuum Breaker 9 Step 4 Install the WH 2600 Cabinet 9 Step 5 Install the Toilet and Waste Connections 10 Step 6 Install the Sationary Lavatory Top 10 Step...

Page 2: ...WH 2600 SERIES PATIENT CARE UNITS INSTALLATION OPERATION MANUAL PAGE 2 Rev 2 2017 1 800 428 4065 TOLL FREE www willoughby ind com PAGE LEFT INTENTIONALLY BLANK...

Page 3: ...ines must be ushed suf ciently to remove the small particles of debris that are inherent with new con struction projects If this debris is not removed the valving in these units will be damaged Do not...

Page 4: ...WH 2600 SERIES PATIENT CARE UNITS INSTALLATION OPERATION MANUAL PAGE 4 Rev 2 2017 1 800 428 4065 TOLL FREE www willoughby ind com PHYSICAL DIMENSIONS WH 2600 SERIES...

Page 5: ...ATIENT CARE UNITS INSTALLATION OPERATION MANUAL PAGE 5 Rev 2 2017 1 800 428 4065 TOLL FREE www willoughby ind com ROUGH IN DIMENSIONS WH 2600 SERIES WH 2600 FLOOR ROUGH IN TEMPLATE WH 2600 BACK WALL R...

Page 6: ...Unit until you obtain the missing parts NOTE Before beginning installation all supply drain and waste piping for the WH 2600 Series must be completed according to speci ed rough ins If you have not r...

Page 7: ...et or cabinet installation if necessary Supply piping to toilet faucet foot valve vacuum breaker bedpan washer AS NEEDED Waste outlet piping to toilet P trap and drain piping Silicone caulk Proper mou...

Page 8: ...ep 2 Install the Anchors and Backing Select the location where the WH 2600 is to be installed Using the supplied templates mark all mounting and rough in locations NOTE Lay out anchor locations for th...

Page 9: ...e NOTE Mounting hardware by others Step 3 Install the Vacuum Breaker for the WH 2600 Series Patient Care Unit w Bedpan Washer 1 Install the supplied vacuum breaker to the wall per local code NOTE The...

Page 10: ...square the toilet assembly to the back wall 3 Level and support the toilet solidly at the four corners using shims if needed shims supplied by installer 4 Secure the toilet in place with appropriate...

Page 11: ...Installation 1 Connect the ush valve outlet to the water closet NOTE Tighten all ush valve connections 2 Connect the ush valve water supply pipe to the ush valve 1 NPT inlet 3 Install Push Button Actu...

Page 12: ...WH 2600 SERIES PATIENT CARE UNITS INSTALLATION OPERATION MANUAL PAGE 12 Rev 2 2017 1 800 428 4065 TOLL FREE www willoughby ind com CARE AND MAINTENANCE...

Page 13: ...ALE REF MAT L TITLE DESC NOTED SIZE EEF 2 PLCS TOLERANCES A Indianapolis Indiana 1 FRACTIONS EEF 0 03 0 015 1 32 Willoughby Industries Inc STANDARD BACK VIEW STANDARD STAIONARY TOILET FOR PATIENT CARE...

Page 14: ...EF 3 5 15 3 5 15 A 1 OF 1 H1490 0079 1 8 N A 2 PLCS SIZE TOILET ASSEMBLY PATIENT CARE 17 TALL SHEET REV ORIG REVISION SCALE REF MAT L PART A ANGLE PETF 1490 FM BS HC 1 6 DC TITLE DESC FRACTIONS 3 PLCS...

Page 15: ...WH 2600 SERIES PATIENT CARE UNITS INSTALLATION OPERATION MANUAL PAGE 15 Rev 2 2017 1 800 428 4065 TOLL FREE www willoughby ind com WARRANTY...

Reviews: