GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS
The Willow Breast Pump Generation 3 is intended for use in the electromagnetic environment specified below.
The customer or user of the Breast Pump should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level Compliance Level
Electromagnetic Environment –Guidance
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
<5 % UT (>95 % dip
in UT) for 0.5 cycle
30% reduction,
25/30 periods at 0°
> 95% reduction, 0.5
period at 0°, 45°,
90°, 135°, 180°, 225°,
270° and 315°
> 95% reduction, 1
period at 0°
> 95% reduction,
250/300 periods
<5 % UT (>95 % dip
in UT) for 0.5 cycle
30% reduction,
25/30 periods at 0°
> 95% reduction, 0.5
period at 0°, 45°,
90°, 135°, 180°, 225°,
270° and 315°
> 95% reduction, 1
period at 0°
> 95% reduction,
250/300 periods
The Willow Breast Pump Generation 3 is suitable
for use in all establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Power frequency (50/60
Hz) magnetic field IEC
61000-4-8
30 A/m
30 A/m
The Willow Breast Pump Generation 3 is suitable
for use in all establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
NOTE UT is the a.c. mains voltage prior to application of the test level.