9. The value “0” indicates that no values are added to the current temperature. A minus in the
display field indicates that a value has been subtracted from the current temperature.
10. Confirm the settings by pressing the up button of the desoldering function
(5).
Example:
An external sensor shows 250
.
℃
The temperature set and displayed on the device is 300
.
℃
300–250 = 50; for the display to be correct again, 50
must be added.
℃
Put the device in calibration mode; the display shows “010.”
Set from “010” to “060” (10 + 50 = 60) by pressing the setting button
(4).
Leave the calibration mode again.
The display should now be correct again.
Fine-adjusting the desoldering gun
The system is calibrated at the factory, but it may be necessary to readjust these settings from time to
time. This can then be done by the following steps:
1. Turn on the desoldering gun.
2. Set the temperature that you would like to calibrate and place the soldering tip on an accurate
external temperature sensor.
3. The results of the external sensor should roughly match those of the system.
4. If there are large differences between an external sensor and the display, the temperature
sensor can be readjusted.
5. Switch off the soldering function on the station
(14);
make sure that the desoldering function is
deactivated
(12).
6. Press and hold the up button of the desoldering function
(4).
7. Release the button when the temperature display of the soldering function
(14)
shows “000.”
You can now calibrate the temperature of the desoldering iron.
8. You can use the temperature adjustment buttons to set the desired value.
9. The value “0” indicates that no values are added to the current temperature. A minus in the
display field indicates that a value has been subtracted from the current temperature.
10. Confirm the settings by pressing the up button of the desoldering function
(5).
Example:
An external sensor shows 300
.
℃
The temperature set and displayed on the device is 350
.
℃
300–350 = –50; for the display to be correct again, it must be decreased by 50
.
℃
Put the device in calibration mode; the display shows “010.”
Set the displayed value from “010” to “040-” (10–50 = –40) by pressing the setting button
(5).
Leave the calibration mode again.
The display should now be correct again.
Maintenance
Spare parts
Name
Item number
Description
10094
90103
Hot air heating element