7-3
USE WITH A 562
SCALAR NETWORK
ANALYZER
The 680XXB is directly compatible with the
WILTRON
Model 562 Sca-
lar Network Analyzer (SNA). The following paragraphs provide in-
structions for connecting the CW generator to the 562 SNA so that is
can be used as a signal source for the analyzer. Operating instructions
for the network analyzer can be found in the Model 562 Scalar Net-
work Analyzer Operation Manual, P/N 10410-00046.
Connecting
the 680XXB
to the 562
Connect the 680XXB CW generator to the 562 sca-
lar network analyzer as shown in Figure 7-2.
Step 1
Connect one end of the Auxiliary I/O ca-
ble (P/N 806-7) to the 562 rear panel
AUX
I/O
connector. Connect the other end of
the cable to the 680XXB rear panel
AUX
I/O
connector.
Step 2
Connect one end of the dedicated system
bus cable (P/N 2100-1) to the 562 rear
panel
DEDICATED GPIB
connector. Con-
nect the other end of the cable to the
680XXB rear panel
IEEE-488 GPIB
con-
nector.
Step 3
Turn on the 680XXB and the 562. The
system is now ready to operate.
USE WITH
USE WITH A 562 SCALAR
OTHER INSTRUMENTS
NETWORK ANALYZER
562 SCALAR NETWORK ANALYZER
680XXB CW GENERATOR
AUX I/O
AUX I/O
IEEE-488GPIB
DUT
DETECTOR
DEDICATED
GPIB
SWR
AUTOTESTER
B
A
Figure 7-2.
680XXB CW Generator to 562 SNA Connections
NOTES
The 680XXB’s GPIB address should
be set to 5 (the default address set-
ting) for operation with a 562 SNA.
To verify or change the GPIB ad-
dress setting refer to Configuring
the GPIB on page 3-62.
The 562 SNA will only accept and
display the nine video markers, F1
thur F9, from the 680XXB.
When performing amplifier testing
only use the 680XXB power level,
L1.
7-10