OPERATION
POWER LEVEL ACCURACY
VERIFICATION
AND FLATNESS TESTS
Model 68095B
Serial No. ____________________
Date ___________
Model 68095B
Power Level Accuracy *
(CW Frequency = 5.0 GHz)
Power Level Accuracy *
(CW Frequency = 25.0 GHz)
Power Level Accuracy *
(CW Frequency = 50.0 GHz)
Set Power
Measured Power
Set Power
Measured Power
Set Power
Measured Power
+10 dBm
+ 9 dBm
+ 8 dBm
+ 7 dBm
+ 6 dBm
+ 5 dBm
+ 4 dBm
+ 3 dBm
+ 2 dBm
+ 1 dBm
+ 0 dBm
– 1 dBm
– 2 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
+ 2.5 dBm
+ 1.5 dBm
+ 0.5 dBm
– 0.5 dBm
– 1.5 dBm
– 2.5 dBm
– 3.5 dBm
– 4.5 dBm
– 5.5 dBm
– 6.5 dBm
– 7.5 dBm
– 8.5 dBm
– 9.5 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
0 dBm
– 1 dBm
– 2 dBm
– 3 dBm
– 4 dBm
– 5 dBm
– 6 dBm
– 7 dBm
– 8 dBm
– 9 dBm
– 10 dBm
– 11 dBm
– 12 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
* Specification is
±
1.0 dB.
* Specification is
±
1.0 dB.
* Specification is
±
1.5 dB.
Power Level Flatness (Step Sweep)
Set Power
Max Power
Min Power
Variation **
–2 dBm
_________dBm
__________dBm
_________dB
** Maximum variation is 1.6 dB (0.5 to 40 GHz); 2.2 dB (40 to 65 GHz).
Table 5-3.
Power Level Accuracy and Flatness Test Record (45 of 46)