Setting
Increment
Sizes
The Increment menu lets you set the incremental
size for editing frequency, power level, and time pa-
rameters. When the increment mode is selected on,
these parameter values will increase or decrease by
the set amount each time the
∧
or
∨
pad is pressed
or the rotary data knob is turned clockwise or
counter-clockwise. The menu also lets you turn the
increment mode on and off.
To go to the Increment menu from the System Con-
figuration menu, press
Incr Menu
. The Increment
Menu (below) is displayed.
Press
Freq Incr
to open the frequency increment
parameter.
Press
Level Incr
to open the power level increment
parameter.
Press
Time Incr
to open the time increment pa-
rameter.
Open the parameter you wish to change, then edit
the current value using the cursor control key or ro-
tary data knob or enter a new value using the key
pad and appropriate termination key. When you
have finished setting the open parameter, close it by
pressing its menu soft-key or by making another
menu selection.
Press
Incr Mode
to turn the increment mode on.
Press again to turn it off.
When done, press
Previous Menu
to return to the
System Configuration menu.