Press
More
to go to the additional Configure GPIB
menu.
Press
Previous Menu
to return to the System Con-
figuration menu.
Additional Configure GPIB Menu
When you press
More
the additional Configure
GPIB menu (below) is displayed.
This menu lets you perform the following:
Select the model and GPIB address for the
power meter that is used to create a user level
flatness correction power-offset table. (Refer to
page 3-55 for a description of the function.)
Select the external interface language for re-
mote operation of 681XXBs with Option 19.
(Refer to page 2-9 for more information.)
Press
Pwr Mtr Address
to change the address of
the power meter on the GPIB (the power meter’s de-
fault GPIB address is 13). Enter a new address, be-
tween 1 and 30, using the cursor control key or the
data entry keypad and the terminator key
Hz
ns
ADRS
The new GPIB address will appear on the display.
Press
Pwr Mtr Select
to select the power meter
model being used. (Supported power meters are the
Anritsu ML4803A and Hewlett-Packard 437B,
438A, and 70100A.)
Press
Native SCPI
to select the external interface
language to be used for remote operation of the
681XXB. (Language selection is only available on
instruments that have Option 19 installed.)
Press
Previous Menu
to return to the main Config-
ure GPIB menu display.
LOCAL (FRONT
SYSTEM
PANEL) OPERATION
CONFIGURATION
681XXB OM
3-73