Setting a Preset Sweep Range–At the menu, se-
lect the sweep range (
F1-F2
,
F3-F4
,
F5-dF
, or
F6-dF
) that you wish to set. The menu then displays
the current frequency parameters for the selected
sweep range. Now, use the menu edit soft-keys to
open the frequency parameters for editing.
Edit the current frequency parameters or enter new
frequency parameter values for the sweep range. To
close the open frequency parameter when you are
finished, press its menu edit soft-key or make an-
other menu selection.
You can set all the preset sweep ranges in this man-
ner.
Selecting a
Power Level
While at the Sweep Frequency Control menu, you
can edit the current output power level or enter a
new output power level for the frequency sweep.
Editing the Current Power Level
Press
Edit L1
to open the power level parameter,
then edit the current power level using the cursor
control key or rotary data knob. To close the open
power level parameter, press
Edit L1
or make an-
other menu selection.
Entering a New Power Level
Press
Edit L1
to open the power level parameter,
then enter the new power level using the keypad
and appropriate terminator key. To close the open
power level parameter, press
Edit L1
or make an-
other menu selection.
NOTE
You can also select any of the preset
power levels for a frequency sweep or a
power level step for analog and step
sweeps. For instructions, refer to the
Fixed Power Level Operation and Power
Level Sweep Operation sections of this
chapter.
LOCAL (FRONT
SWEEP FREQUENCY
PANEL) OPERATION
OPERATION
681XXB OM
3-31