6000Counts Pocket Type Intelligent Scanning Digital Multimeter

6

/

8

4. Capacitance(C)

Overload protection: 1000V DC, 750V AC peak.
4-1. The boot display is in the automatic scan status "AUTO".
4-2. The black test lead is negative, the red test lead is positive, and the test lead reliably touches the test point.
4-3. If you need fast continuity test/diode/capacitance measurement, constantly trigger the "power" switch, enter the fast continuity

test/diode/capacitance measurement in one cycle, and select the corresponding function measurement according to the measurement requirements.
When measuring the capacitance, the measured capacitance The size will automatically select different ranges, and the measured value will be
displayed on the LCD. The capacitance measurement range is 10nF/100nF /1uF/10uF/100uF/1mF/10mF/60mF.

Note:
1) When measuring capacitance with 10nF file, there may be residual reading on the screen display value. This number is the distributed capacitance

of the test pen. It is an accurate reading and can be subtracted after measurement.

2) When the large capacitance file measures severe leakage or breakdown capacitance, some values will be displayed and unstable; when measuring

large capacitance, the reading takes several seconds to stabilize, which is normal when measuring large capacitance;

3) Please fully discharge the capacitor before testing the capacitor capacity, otherwise it will enter the voltage measurement mode.
4) Unit: 1F=1000mF, 1mF=1000uF, 1uF=1000nF, 1nF=1000pF

Accuracy

Range

6000counts

Resolution

10nF

±(

3.5%+20

10pF

100nF

100pF

1uF

1nF

10uF

10nF

100uF

100nF

1mF

1uF

10mF

10uF

60mF

±(

5%+3

100uF

Summary of Contents for ET8138

Page 1: ...of Pocket Type Intelligent Scanning Digital Multimeter Index 1 Overview 1 2 Safety precautions 1 3 Characteristics 1 4 Operation panel instructions 3 5 Technical Specifications 4 6 Automatic switch ma...

Page 2: ...superior tool and an ideal tool for laboratories factories radio enthusiasts and families 2 Safety Precautions This series of instruments is designed in accordance with IEC1010 Safety Standards promul...

Page 3: ...haracteristics 2 1 Accuracy a of reading least significant digit to ensure accuracy Ambient temperature 23 5 relative humidity 75 calibration guarantee period from the date of manufacture for one year...

Page 4: ...frequency and buzzer Black is the negative end of the input Note 1 The back light is always on LCD is highlighted and the user is constantly on In use the instrument will not turn off the power supply...

Page 5: ...the red test lead is greater than 0 8V regardless of the AC voltage or the DC voltage the meter will compare the DC component and the AC component take the larger component signal and then according...

Page 6: ...if the voltage in the loop is greater than 0 8V the meter will mistake it for voltage measurement and enter the electricity measurement mode 2 5 When measuring the value at the negative end of the bl...

Page 7: ...ill be displayed on the LCD The capacitance measurement range is 10nF 100nF 1uF 10uF 100uF 1mF 10mF 60mF Note 1 When measuring capacitance with 10nF file there may be residual reading on the screen di...

Page 8: ...e contact with the point the black meter end is suspended and does not contact any test point if the test point is live wire the LCD will display OL and the buzzer will send continuous sound 6 Frequen...

Page 9: ...completely welded at the input end there will be certain data on the LCD In the design we define the manual DC voltage level the input end has 10 words the AC voltage level has 100 words and the capac...

Reviews: