OWNERS OPERATING & MAINTENANCE MANUAL
TMM3 VIDEO FLUOROSCOPY STRETCHER-CHAIR
Page 5 of 28
Document #OM-TMM-2228-13
Revision: V
POWER / CABLE REQUIREMENTS
For options A, B & E, see the "Options" section of this manual for additional details regarding powering
this product.
Only use the power/charging cord that came with your TMM3 Video Fluoroscopy Stretcher-Chair.
SAFETY PRECAUTIONS
WARNINGS
CHAIR OPERATION BY QUALIFIED, TRAINED MEDICAL PERSONNEL ONLY
The chair is intended to be operated only by qualified, trained medical staff.
Operation of chair by unauthorized, untrained and/or lay people must be avoided.
USE CAUTION ON RAMPS
Control chair when traversing ramps. If a collision occurs, serious injury to patient,
bystanders, or medical personnel and damage to chair or medical facility could
occur.
LOCK CASTERS BEFORE PATIENT EGRESS AND INGRESS
Prior to patient egress and ingress, casters must be locked by pressing the red
end down on any of the brake pedals located on the base of the chair.
AVOID PINCH POINTS AND OTHER INJURIES
To prevent serious injury, ensure extremities of patient and bystanders are clear
of all mechanical systems when operating motors for lift and positioning functions.
To prevent pinch / crush injury, ensure extremities of patient and bystanders are
clear of locking mechanism when raising and lowering side rails.
To prevent patient strangulation, use approved hand pendant storage location
when not in use. See PENDANT section for additional information.
BATTERY CAN EXPLODE DUE TO OFF-GASSING WHEN CHARGING
At end of charging process (or with overcharge conditions), battery can produce
mixture of explosive gases (including hydrogen and oxygen). Avoid exposing
battery to open flames, cigarettes, sparks, and incandescent materials.
Never charge battery in enclosed, unventilated spaces.
Do NOT store battery in sealed container. Store in fresh, well-ventilated area
protected from direct sunlight and heat sources.
Do NOT use water to extinguish battery fire. Use dry powder, foam CO2
extinguisher.
USE FACTORY APPROVED BATTERIES ONLY
Use of non-factory approved batteries will void all warranties and could result in
unsatisfactory performance and/or damage.