OWNERS OPERATING & MAINTENANCE MANUAL
TMM3 VIDEO FLUOROSCOPY STRETCHER-CHAIR
Page 22 of 28
Document #OM-TMM-2228-13
Revision: V
SWITCH SCHEMATIC
•
"A" Option Only: Ground wire attaches to
GROUND POST
•
Battery cord plugs into
PORT #13
•
Plug column cord plug into
phono din adapter, then plug
into
PORT #1
•
PORT #2
is empty
•
Back section
actuator plugs
into
PORT #3
•
Leg section actuator plugs into
PORT #4
•
PORT #5
is empty
•
Seat rotation cord plugs into
PORT #7
•
Pendant cable plugs into
PORT #9
•
All excess wire should be placed in FRONT of the Control Box.
CLEANING INFORMATION
STEAM AND/OR PRESSURE CLEANING CHAIR WILL VOID WARRANTY
TMM3 STRETCHER-CHAIR
Wipe TMM3 Stretcher-Chair (rails, base cover, etc.) with a damp cloth and allow to air dry.
NOTE: Clean exterior surfaces of electronics (actuators, battery pack, pendants) with minimal
water.
Complete above steps as required and/or in accordance with facility policies.
A NOTE ABOUT CLEANING AGENTS
Facility approved detergents, disinfectants and water are suggested.
Use of facility approved detergents and disinfectants must comply with the
instructions provided by the manufacturer(s) of those products.