Medical Windows Tablet M101P-ME User Guide
5.2 Replacing Battery
The M101P-ME Tablet supports hot-swap, meaning the tablet doesn
’t need to be powered off to
replace the main battery. To support hotswap, there is a backup battery inside the tablet that is charged
together with the main battery. This battery cannot be removed, and requires adequate power for main
battery hot swapping. When the backup battery is fully charged, the tablet can sustain up to 5 minutes
of runtime for battery hot-swapping.
Perform the following to hot swap the batteries:
1. Power on the tablet, open Hottab to check the power level of the Backup battery.
2. If the Backup battery level is
“
Low
”, please charge the tablet with AC Power until the Backup
battery level is
“
Medium
” or “
High
”, and do not remove the main battery.
3. If the Backup battery level is at
“
Medium
” or “
High
”, you can remove the main battery.
4. Insert a new fully charged battery into the battery compartment.
5. Pen Hottab to check Main battery level. If the battery level is displayed incorrect, check if the
new removable battery is inserted properly.
6. If the battery level displayed correct, continue using the tablet.
Caution
:
Charge the built-in back-up battery every two months
to
keep battery
health and prevent possible issues from battery over-discharge.
Attention:
chargez la batterie de secours intégrée tous les deux mois pour
préserver son intégrité et éviter les problèmes de décharge excessive de la batterie.
5.3 Using Recovery Wizard to Restore Computer
The M101P-ME Medical Tablet Computer has a dedicate recovery partition stored on the hard drive of
the tablet to enable quick one-key recovery process. This partition occupies about 11GB of the storage
space, and comes built-in to each M101P-ME tablet computer.
Warning!
Before starting the recovery process, be sure to backup all user data, as all data will
be lost after the recovery process.
To enable quick one-key recovery procedure:
1. Plug-in the AC adapter to the tablet computer. Make sure the tablet computer stays plugged in
to power source during the recovery process.