Product: Eleganza 1
Version: 01
29/68
The manufacturer recommends to replace the battery by qualified service organization after 2 (two) years
of use. After this period the supposed service life of battery ends and the manufacturer cannot guarantee
the battery service life after this period.
The battery must be replaced with the new battery approved by manufacturer after maximum 5 (five)
years of use at the latest.
The manufacturer will assume no responsibility for any damage to the bed or the battery caused by:
o
Non-
observance of the manufacturer’s instructions in the user manual.
o
Using batteries not approved by the manufacturer.
o
Battery replacement non-qualified service organisation.
To charge the battery:
Connect the bed to the mains and check the yellow LED on the Supervisor control panel according to the
table 1.
Charging and battery capacity is indicated by the yellow LED placed on the ACP control panel.
NOTE: Some bed adjustment options are not available without a battery, for example, height adjustment under a
load of above 200 kg.
The LED indicates the battery’s charge status:
Yellow LED
Battery charge status
Not lit
Battery capacity is sufficient (charging completed)
Short flashing (short, intermittent illumination)
(circa 1.8 sec.)
Battery is charging - continue charging until the LED
is extinguished. In emergency cases, the battery can
be used as a backup power source for a short period.
If LED is still flashing after 12 hours of charging or
stops flashing, but you cannot position with bed,
battery is defective or broken. Contact manufacturer.
Long flashing (long illumination) (circa 0.2 sec.)
Low battery voltage - battery cannot be used as a
backup power supply even for a short period; battery
is completely discharged or defective (if this type of
signalisation persists, it is necessary to replace the
battery - service action)
Long, intermittent illumination for several hours (circa
10 hours), when bed is connected to the mains.
Battery absence or failure condition (battery is
connected incorrectly, line between the power supply
and battery is broken or battery fuses are faulty);
contact service department of the manufacturer in
case of such signalisation.
Table 1
ACP control panel
– battery signalization
To maintain maximum functionality of the battery:
Unplug the bed from the mains as least as possible.
In case the battery cover or control section is deformated by heat
Unplug the bed from the mains.
Do not use the bed (see. Removing the Bed from Service.
Contact service of the manufacturer.