10.2.5 Position Buttons
The therapeutic and safety-related positions are pre-programmed. When a position is set, several parts of the bed
and mattress platform will move simultaneously.
Programmed positions:
Cardiac chair position
13
Trendelenburg position
12
CPR (resuscitation) position
11
To set the programmed positions:
Activate the keypad by pressing the GO button.
Press and hold corresponding function button until the desired position is reached.
Cardiac chair position (button 13)
The cardiac chair position is suitable for patients with cardiac arrhythmia and breathing difficulties.
Settings after pressing and holding the Cardiac chair button (13):
Calf rest tilts into lowest position, thigh
rest moves into the upright position (34°).
6 second after positioning calf rest and thigh
rest the backrest will move into an upright position (62°).
Trendelenburg position (12)
The Trendelenburg position serves as an anti-shock position. All parts of the mattress platform are flattened.
Mattress platform tilts head down.
CPR (Resuscitation) Position
The CPR position is for resuscitating the patient in an emergency.
Settings after pressing and holding the CPR button (11):
Mattress platform moves into a horizontal position.
NOTE: For quick mechanical positioning, see CPR Backrest Release.
WARNING!
Risk of injury due to moving parts!
Ensure that no body parts are trapped between moving parts of bed and mattress platform.
Ensure that no persons or body parts are close to bed or accessories (e.g. infusion stand, lifting pole) when
mattress platform is moving
CAUTION!
Damage to property due to moving parts!
Ensure that no objects (e.g. cables) are trapped between moving parts of bed and mattress platform.
Ensure that no objects are close to bed or accessories (e.g. infusion stand, lifting pole) when mattress
platform is moving.