Product:
Latera
Version: 01
19
/57
Yellow LED
Battery status
Not lit
Battery capacity is sufficient (charging completed)
Short flashing (shortly lit,
longer not lit) (circa 1.8 sec.
Battery is charging - continue charging until the LED is extinguished. In
emergency cases, the battery can be used as a backup power source for a
short period. If LED is still flashing after 12 hours of charging or stops flashing,
but you can not position with bed, battery is defective or broken. Contact
manufacturer.
Long flashing (longer lit, shortly
not lit) (circa 0.2 sec.))
Low battery voltage - battery can not be used as a backup power supply even
for a short period; battery is completely discharged or defective (if this type of
signalisation persists, it is necessary to replace the battery - service action)
Lit continuously for several
hours (circa 10 hours), when
bed is connected to the mains.
The LED indicates the
battery’s charge status:
Battery absence or failure condition (battery is connected incorrectly, line
between the power supply and battery is broken or battery fuses are faulty);
contact service department of the manufacturer in case of such signalisation.
Replacing the battery
Damage to the bed due to incorrect battery replacement!
Have the battery replaced exclusively by qualified personnel.
Exclusively use batteries approved by the manufacturer.
Caution
Warning
Risk of damage or destruction of battery!
If battery is faulty degassing can occur. In rare cases this can cause deformation of
the battery pack housing and mains control box.
If this occurs the bed must be immediately taken out of service and taken to an ade-
quately ventilated room without sparks (electricity or fire)!
Immediately inform service department of the manufacturer!
Warning
Danger of reducing battery durability due to wrong use!
Use bed on battery only in crisis situations (e.g.: power blackout, patient complicati-
ons during transport, etc.)
After reconnecting bed to the mains charge battery to full capacity (see chart Battery
charge status
Battery must be replaced with new battery approved by the manufacturer.
Bed must only be fitted with batteries approved by the manufacturer. To get more information on how to
change battery please refer to the Service manual (contact service department of the manufacturer).