109
ENGLISH
Alarm is triggered without any
apparent reason
Nursecall system adapter is
defective
Contact in-house technician,
otherwise contact the WiBo
customer service
Cable is defective
Sensors are defective
Missing magnets
Weight is not displayed correctly
or values of the weight history
differ too much
Weight was added/removed
without freezing it
Ignore extremely differing/
wrong values
Weight was set to zero acci
-
dentally
Wrong measuring
Display shows „hi“ or „lo“
Zero point set incorrectly
Set the display to zero when the
resident is not in bed
Underbed light is not working as
it should
Wrong standard saved (patient
is in the bed when turning on
automatic functions)
Switching underbed light into
automatic mode when the
resident is in bed
Handset does not work
Handset locked, not correctly
plugged in or faulty
Unlock handset, check connec
-
tion. Otherwise contact the
customer service
Mobilift
®
does not work
GO button is not pressed
Press the GO button (handset or
CareBoard
®
)
Handset is locked
Unlock handset
Mobilift is defective (cable
break)
Contact the WiBo customer
service
CareBoard
®
does not react
properly
You are pressing the touch
screen too hard or the area
being pressed is too large
When operating the board, plea
-
se make sure that you press the
center of the button and do not
push too hard
The power box emits an
acoustic signal
CareBoard
®
not correctly
connected / faulty
Check the plug connection or
call WiBo service
Save & Send button lights up
red
Fault affecting the server
connection
Check network cable, reset the
CMC box (see chapter 5.2) – or
contact WiBo service