80
ENGLISH
5.11 Height adjustment
Bed egress position
When adjusting the height of the bed upward or downward, the bed stops at the comfort bed
egress position. To continue adjusting the bed, the function key on the handset or CareBoard
®
needs to be released and then pressed again. Using movable lifting equipment
(e. g. mobile patient lift)
The low nursing bed´s design requires a safety clearance of less than 150 mm.
Make sure before each use of movable lifting equipment that the free space be
-
neath the low nursing bed is sufficient for the movable lifting equipment. In any
other case, adjust the height before using movable lifting equipment.
Attention!
Comfort seating position / legs up position: Particularly in conjunction with a bed
extension, there is a danger of entrapment between the extension and the floor –
and of the foot section or head section coming into contact with the floor. Move the
patient surface to an appropriate height before every adjustment of the bed. The
same applies when a linen holder is used!
Danger!
Before every height adjustment of the bed, you must make sure that there are no
persons, animals or objects in the danger area. Always leave the handset in the
locked position so that no unauthorized persons can manipulate the height adjust
-
ment of the bed.
Attention!
Due to its design as a low nursing bed, there is a clearance of less than 150 mm.
Prior to every height adjustment of the bed, you must make sure that there are no
persons, objects or anything else in the entrapment area beneath the bed! Always
leave the handset in the locked position so that no unauthorized persons may ad
-
just the height of the bed.
Attention!
When lowering the bed or tilting it to the seating position, make sure not to get
your feet in the danger area (at the bed ends or under the bed).
Summary of Contents for sentida 7-i
Page 350: ...350...