3.3.6 

The menu «Delete log data» 

Be careful – using this function will delete all measurement data! 

 

Select «Delete log data» and confirm with 

 

To delete all stored measurement, confirm the following security query with “YES”. 

 

3.3.7 

The menu item «Back» 

Menu item to leave the menu without performing any (further) action / setting. 

 

3.4 

The menu «Device Settings» 

 

 

Select (

 

) «Device Settings» in the main menu and confirm with 

 

 

 

3.4.1 

The menu «Date / Time» 

 

Select «Time / Date» and confirm with 

 

 

Select «Date» or «Time» in the submenu and confirm with 

 

Set as described in Section 3.1.1. 

 

 

3.4.2 

The menu «Languages» 

 

Select the desired language. 

 

Pressing 

 confirms the selected language. 

 

To exit the menu, select “Back” and confirm with 

 

3.4.3 

The menu item «Device info» 

After selecting this menu item, the software version and serial number will be indicated. 

 

3.5 

The menu item «NFC enable» 

After selecting this menu item, the device will try to establish a 

N

ear 

F

ield 

C

om-

munication to an external device (e.g. a smart phone as used in the following 
descriptions).  
 

 

Start the application «PATBOX control» at the smartphone. 

 

Place the back of the PATBOX to the back of the smartphone.  

Note:

 To get a stable connection, the PATBOX should be placed directly above the NFC an-

tenna of the smartphone. If their position is unknown, move the PATBOX until a stable connec-
tion is established. 

After establishing a connection there is an acoustic signal of the smartphone. In 
addition, the buttons for triggering the PATBOX are not shown greyed out any 
more. In addition, “connect” is indicated on the display of the PATBOX. 

You find references for the application «PATBOX control» in Section 5. 

 

To stop this function press 

 

Time / Date

  

Languages  

NFC Enable  

Device info 

Back 

Date settings 

          day   month  year 

        19 . 03 . 

2018

   

                           ok 

Time settings 

          hour   min   sec 

        11 : 

03

 : 28   

                           ok 

 

 

connected 

Summary of Contents for PATBOX

Page 1: ...Copyright WITT GASETECHNIK GmbH Co KG 2018 all rights reserved Instruction Manual for WITT PATBOX Issue 07 06 2018 This issue is not subject to change management...

Page 2: ...logging 5 3 3 2 The menu Start type 6 3 3 3 The menu Interval 6 3 3 4 The menu Countdown 6 3 3 5 The menu Show log data 6 3 3 6 The menu Delete log data 7 3 3 7 The menu item Back 7 3 4 The menu Devi...

Page 3: ...ement There are several methods to start a measurement Start on pressing a button start after a settable count down time expired or start on movement of the PATBOX By means of a NFC connection log dat...

Page 4: ...he actual measured temperature The actual measured pressure The indicated measurement values are not stored With button you proceed to the main menu With buttons you proceed to the indication of store...

Page 5: ...art type Section 3 3 2 During logging the display will be deactivated 3 2 1 Check stop a running logging To check weather a logging is in process press or If this is the case Logging will be indicated...

Page 6: ...menu Interval In this menu you can select the time interval min 100 milliseconds max 24 hours for storing measurement values Set the desired time interval in hours minutes seconds 100 milliseconds as...

Page 7: ...em Device info After selecting this menu item the software version and serial number will be indicated 3 5 The menu item NFC enable After selecting this menu item the device will try to establish a Ne...

Page 8: ...charging The charging indication will be indicated some time after putting the PATBOX on the charging pad The date and time setting can be lost After charging these settings should be checked and corr...

Page 9: ...m Show Chart Select the serial number of the desired PATBOX if there are data of more than one device Select the desire data for indication Note If you select the desired data for about 2 seconds you...

Page 10: ...Logging interval Settable from 0 1 second to 24 hours in steps of 0 1 s Memory Flash chip Capacity 8 MB Number of records approx 600 000 Communication NFC Near Field Communication Housing Shock resis...

Page 11: ...must accept any interference received including interference that may cause un desired operation NOTE Changes or modifications made to this equipment not expressly approved by manufacturer name may v...

Reviews: