background image

 

While  looking  through  the  eyepiece  and  using  the  course-focus 
adjustment,  slowly  move  the  stage/specimen  AWAY  from  the 
objective until the image is clear. 

 

Rotate  the  10X  objective  into  place.  While  looking  through  the 
eyepiece  and  using  the  fine-focus  adjustment,  clearly  focus  the 
image,  that  now  appears  enlarged  compared  to  that  observed 
through  the  4X  scanning  objective.  You  will  notice,  increasing 
magnification  decreases  the  field-of-view  or  relative  observable 
area of the specimen. 

 

At this point, it may be beneficial to diminish the amount of light 
entering  the  specimen.  Decreasing  the  aperture  of  the  iris 
diaphragm  will  reduce  glare  and  increase  contrast  in  the  image. 
Use of the iris diaphragm will require experience. It is important to 
select  both  the  magnification  and  the  degree  of  contrast  that 
renders  the  optimum  detail  in  the  observed  specimen.  You  may 
also need to regulate the amount of light with the Dimmer Dial. 

 

Rotate the 40X objective into place and repeat the steps taken for 
the initial magnification increase. The objectives of the microscope 
are  parfocal,  meaning  that  the  gross  image  will  remain  in  focus 
when changing through the objectives. It should only be necessary 
to  use  the fine-focus  adjustment  once  the  initial  image  is  clearly 
viewed  at  low  magnification.  With  the  40X  it  is  necessary  to  be 
careful not to touch the slide or any liquid specimen. The 40X is 
not sealed against liquid and could be damaged. If the specimen 
will not focus, make sure that you are using the correct thickness 
of slides.  

 

In order to obtain maximum resolution of the 100x oil immersion 
lens, it is necessary to apply immersion oil to the top of the cover 
glass on the slide. Use the smallest amount possible. A tiny drop 
goes a long way in use. It is extremely important that the oil from 
the  100x  is  not  transferred  to  the  other  objectives.  The  oil  will 
damage the non-oil immersion lens. 
1.

 

Use a very small amount of Immersion oil. Immersion Oil can 
be purchased from Carolina. # 853311 15ml bottle. 

2.

 

After viewing with the 100 x lens, clean the oil from the glass 
surface  and  anywhere  the  oil  has  come  in  contact  with  the 
objective. It is important to clean all the oil from the slide as 
well so that the oil does not damage the other objectives. 

 

It  will  be  useful  to  repeat  these  procedures  until  they  become 
second  nature.  Your  familiarity  with  them  will  add  to  the 
usefulness  of  the  microscope  and  to  the  ease  in  interpreting  the 
information it provides. 

 
 

Reviews: