be  cleaned  with  optical/lens  tissue  or  pure  cotton.  For  stubborn  fingerprints  or 
smudges, breathe against the lens surface and wipe gently with lens tissue. AVOID 
THE USE OF SOLVENTS OR CLEANING SOLUTIONS AS THESE AGENTS MAY DAMAGE 
LENS ELEMENTS THAT HAVE BEEN CEMENTED IN PLACE. The 4x, 10x and 40x are 
made to be used with dry slides only. Do not allow the objectives to touch a wet 
specimen or touch oil. This could damage the objectives and is not covered under 
the  warranty.  Always use  a  coverslip  on  slides  when using  the  10  x  and  40  X  to 
protect the lens.  
The  100  X  Oil  Immersion  Objective  will  not  focus  on  a  dry  slide.  You  must  use 
Immersion Oil with this lens to focus the image. The specimen must have a cover 
clip to protect the specimen and lens. With the tiniest drop of oil on the coverslip, 
bring the 100 X down very carefully to the slide. Allow the oil to make contact with 
the lens. You may at this time need to adjust the Iris Diaphragm and Dimmer Dial 
to allow more light into the objective. Now you can do some critical focusing with 
Coarse  and  Fine  Focusing  Knobs.  IT  IS  EXTREMELY  IMPORTANT  NOT  TO  RETURN 
THE  NOSEPIECE  TO  THE  40  X  OBJECTIVE  AFTER  USING  IMMERSION  OIL.  After  all 
viewing with the 100X is complete. It is extremely important to lower the stage all 
the way down and clean all the oil from the 100 X and the slide. Any oil that comes 
in contact with the other objectives can damage them.   
MECHANICAL  STAGE:  The  Beta  Elite  microscope  is  equipped  with  an  integral 
mechanical  stage  and  the  Specimen  Holder  is  installed  at  the  factory.  To  load  a 
slide onto the stage, simply bring the slide to the holder from the right-hand side 
of the holder and place onto the stage so that the  spring-loaded clip on the left 
side will secure the slide. To remove the slide, reverse this action by picking up the 
slide in the middle or end and push the left clip outward and pull the slide up. 
FOCUS CONTROL (stage type): The Wolfe

 Beta Elite Series microscopes utilizes a 

fixed body tube. Focus is achieved through use of coaxial course and fine-focus 
adjustment  knobs  located  on  the  lower  half  of  the  microscope  arm  and  block, 
below the stage. You should notice vertical movement of the stage when focusing 
an image. NOTE: On the left side of the microscope, the inner ring on the coaxial 
focus knob is a tension adjustment ring. Rotation of the collar away from the stage 
increases tension of the focus adjustment. Rotation of the ring 
BUILT-IN LED ILLUMINATOR: Your microscope has a built-in LED illuminator. AS A 
PRECAUTION,  GROUNDED  ELECTRICAL  SERVICE  SHOULD  ALWAYS  BE  PROVIDED 
WHEN  USING  THE  MICROSCOPE.  The  LED  is  rated  at  40,000  hours,  so  it  is  likely 
you  will  not  need  to  replace  them  for  many  years.  In  the  event  a  new  bulb  is 
needed, you can order item # 591141 

Wolfe® Beta Elite LED Replacement Lamp.

 This 

bulb is a lamp assembly and will need to be installed inside the scope. If you are 
not  sure  about  opening  your  scope,  contact  the  microscope  department  at 
Carolina for instructions.  