 and fit the emergency release cap back in the hole.

11.5 IMBALANCE DETECTION

This centrifuge is equipped with an imbalance detec on safety feature. When the 
rotor is not loaded symmetrically, imbalance detector gets ac vated. This will cut 
off the centrifuga on & error “Err55” will be shown on the display. To correct the 
imbalance, open the centrifuge lid and correct the imbalance load using methods 
described in the balancing the rotor sec on (page no.10) of this manual. A er 
correctly loading the tubes into the rotor turn off the centrifuge & turn it back on.

12. SAFETY PRECAUTION 

1.  Never use the centrifuge in any manner not specified in the manual.
2.  Never move the centrifuge while it’s opera ng.
3.  The  rotor  must  always  be  securely  fastened.  If  the  centrifuge  makes  an 

unusual noise during the opera on, the rotor fit needs to be checked.

4.  The rotor must be loaded symmetrically.
5.  Do  not  use  the  centrifuge  or  rotor,  if  it  is  not  installed  correctly  or  it  is 

damaged.

6.  Do not use the centrifuge or rotor if it shows any sign of external mechanical 

or electrical damages.

7.  The centrifuge must be used only for the specified applica on. It must not be 

operated  in  a  hazardous  or  flammable  environment  &  no  hazardous  or 
flammable content should be centrifuged.

8.  Do not fill tubes while they are in the rotor. Liquid spillage may harm the 

device.

9.  If  liquid  is  spilled  on  the  rotor  or  rotor  chamber,  the  centrifuge  must  be 

cleaned immediately.

10.  “Tube Holders” are a part of the Rotor. They should be placed in the rotor 

whether 2,3,4,6 or 8 tubes are being used.

13. TROUBLESHOOTING  

This  centrifuge  has  a  self  –  diagnos c  func on.  If  a  problem  occurs,  an 
error/warning code will be displayed on the display screen and the operator can 
determine the malfunc on with the warning code below.

ERROR

PROBLEM

SOLUTION

No display

No main power 
connection.

Check power & properly plugin power 
adaptor at both ends.  

Power failure

Check the main fuse of the lab.

Improper connection.

Connect adaptor properly.

9

Summary of Contents for Clinispin CT25

Page 1: ...Clinispin CT25 User Manual...

Page 2: ......

Page 3: ...1 4 TechnicalSpeci ca on 1 5 StandardAccessories 2 6 Installa on 2 7 StandardPartsLis ng 3 8 TubeHolder Adaptors 4 9 RotorInstalla on 6 10 BalancingtheRotor 6 11 Opera on 7 12 SafetyPrecau on 9 13 Tr...

Page 4: ...ancedetectorwithautocut o forsafeopera on Brushless DC motorformaintenancefreeperformance Lidlocksafety thelidopensautoma callyonruncomple on Digitaldisplaywithspeedand meindicator RPM se ngfrom500 65...

Page 5: ...the packaging and accessories are removed from the rotor chamber The rotor is pre installed Pleasemakesurethatrotoris rmly ghtened Accessoriesanduser manual are kept inside the box with the centrifug...

Page 6: ...lease match the s cker posi on as showninthe gurebelow gentlypushitin 4 If matched correctly the plug will go in easily without much force do not try toforcethepluginifitisnotmatchedproperly Oncethepl...

Page 7: ...ps three di erent reduced adaptor can be designed to hold di erent types of blood collec ontubes The di erent adaptor combina ons along with tube holder for di erent blood collec ontubesareillustrated...

Page 8: ...R C The maximum diameter and proposed length for such combina onare MaximumTubeDiameter 13mm ProposedTubeLength 55to75mm Check both the criteria Max Diameter Proposed Length of the blood collec on tub...

Page 9: ...clean Load the rotorver callyintothemotorsha 2 PuttheT Allenkeyintherotornut turnclockwiseto ghtentherotor 3 A erproperlyfasteningtherotor placetheTubeHoldersintotherotor 10 BALANCING THE ROTOR 1 Alwa...

Page 10: ...operly installing the centrifuge switch ON the centrifuge from the rear sidebypressingthepowerswitch Thelidstatuswillbeseeninthedisplay Ifthelid isopen thedisplaywillshow Lid andifitisclosed theset RP...

Page 11: ...e RPM RCF to see corresponding RCF value 11 3 START STOP OPERATION Once all the parameters like Speed Time are set press the Start bu on to START opera on 1 Makesurethatthelidisclosedbeforestar ngthec...

Page 12: ...r rotor if it is not installed correctly or it is damaged 6 Do not use the centrifuge or rotor if it shows any sign of external mechanical orelectricaldamages 7 The centrifuge must be used only for th...

Page 13: ...te all liabili es on us for the products or damages caused by its use The buyer and its customer shall be responsible for the product or use of products as well as any supervision required for safety...

Page 14: ...amage We will not be responsible for damage incurred byimproperpackaging All items returned for service should be set postage prepaid in the original packaging or other suitable carton added to avoid...

Page 15: ......

Page 16: ...ngdom Tel 44 0 8456 777001 Fax 44 0 8456 777002 enquiries woodleyequipment com Woodley Equipment Company Inc USA O ce 37 18 Northern Boulevard STE 512 Long Island City NY 11101 United States Toll Free...

Reviews: