and fit the emergency release cap back in the hole.
11.5 IMBALANCE DETECTION
This centrifuge is equipped with an imbalance detec on safety feature. When the
rotor is not loaded symmetrically, imbalance detector gets ac vated. This will cut
off the centrifuga on & error “Err55” will be shown on the display. To correct the
imbalance, open the centrifuge lid and correct the imbalance load using methods
described in the balancing the rotor sec on (page no.10) of this manual. A er
correctly loading the tubes into the rotor turn off the centrifuge & turn it back on.
12. SAFETY PRECAUTION
1. Never use the centrifuge in any manner not specified in the manual.
2. Never move the centrifuge while it’s opera ng.
3. The rotor must always be securely fastened. If the centrifuge makes an
unusual noise during the opera on, the rotor fit needs to be checked.
4. The rotor must be loaded symmetrically.
5. Do not use the centrifuge or rotor, if it is not installed correctly or it is
damaged.
6. Do not use the centrifuge or rotor if it shows any sign of external mechanical
or electrical damages.
7. The centrifuge must be used only for the specified applica on. It must not be
operated in a hazardous or flammable environment & no hazardous or
flammable content should be centrifuged.
8. Do not fill tubes while they are in the rotor. Liquid spillage may harm the
device.
9. If liquid is spilled on the rotor or rotor chamber, the centrifuge must be
cleaned immediately.
10. “Tube Holders” are a part of the Rotor. They should be placed in the rotor
whether 2,3,4,6 or 8 tubes are being used.
13. TROUBLESHOOTING
This centrifuge has a self – diagnos c func on. If a problem occurs, an
error/warning code will be displayed on the display screen and the operator can
determine the malfunc on with the warning code below.
ERROR
PROBLEM
SOLUTION
No display
No main power
connection.
Check power & properly plugin power
adaptor at both ends.
Power failure
Check the main fuse of the lab.
Improper connection.
Connect adaptor properly.