IMPORTANT NOTE:
1.
If the centrifuge is working slow or hot due to overdraw current, restart the centrifuge and
check it again.
2.
Do not use liquid with density higher than 1.2g/ml for full load opera on
14. WARRANTY STATEMENT
This product is warranted to be free from defects in material and workmanship for
a period of two (2) years from date of purchase. Your product will be duly repaired
upon prompt no fica on in compliance with the following condi ons:
This warranty is valid only if the product is used for its intended purpose and
within the guidelines specified in this instruc on manual. This warranty does not
cover damage caused by accident, neglect, misuse, improper service, natural
forces or other causes not arising from defects in original material or
workmanship. This warranty does not cover any incidental or consequen al
damages, commercial loss or any other damages from the use of this product.
The warranty is invalidated by any non-factory modifica on, which will
immediately terminate all liabili es on us for the products or damages caused by
its use. The buyer and its customer shall be responsible for the product or use of
products as well as any supervision required for safety. If requested the products
must be returned to the distributor in well packed and insured manner and all
shipping charges must be paid. Some states do not allow limita on on the length
of implied warran es or the exclusion or limita on of incidental or consequen al
Err 55
Rotor not loaded
symmetrically.
Load the rotor symmetrically
& restart the centrifuge.
Centrifuge lid
cannot be
opened
Rotor is still spinning.
Wait for the rotor to come to a stop.
Power failure
Press emergency lid release after
rotor stops
Centrifuge
shakes during
acceleration
& makes loud
running
noises while
running
Rotor is not loaded
symmetrically.
Load rotor symmetrically & restart
operation
Either a broken tube,
damage to the rotor or
motor is the cause for
run noise.
Replace broken tube. For damaged
rotor/motor contact a service
representative.
Rotor is damaged.
Remove & change rotor
Err 52
Rotor is Stuck
Turn OFF the centrifuge, Check & fit the
rotor properly & turn ON the centrifuge
10