Err 2 

Lid is not properly latched 

Open lid and close it 

properly 

Err 52 

Rotor is stuck 

Turn the centrifuge off. 

Check and fit the rotor 

properly and turn the 

centrifuge on 

Power tripping 

Cable does not fit 

properly 

Remove cable and 

connect properly 

Last run memory not 

displayed 

The centrifuge was 

turned ON immediately 

after being turned OFF 

Maintain a 3 second gap 

between turning the 

centrifuge OFF and ON 

System hang/freezes 

Electronics error 

Switch off centrifuge and 

then switch it ON again. If 

the error still shows, 

contact Woodley 

Equipment Company 

 

NOTE

:

  

1.

 

Maintain  a  3  second  gap  between  restarting  the  centrifuge.  Instantly 
switching ON-OFF can lead to a reset, erasing the last run memory.

 

2.

 

If the centrifuge is hot which may result in a fluctuation in speed then allow 
it to cool for at least 30 minutes whilst not using it. 

3.

 

Do not use liquids with density higher than 1.2g/ml for full load operation.

 

 

14.

 

 Warranty 

This product is warranted to be free from defects in material and workmanship for a 
period of two (2) years from date of purchase. Your product will be duly repaired 
upon prompt notification in compliance with the following conditions: 

This warranty is valid only if the product is used for its intended purpose and within 
the guidelines specified in this manual. This warranty does not cover damage caused 
by accidents, neglect, misuse, improper service, natural forces or other causes not 
arising from defects in original material or workmanship. This warranty does not 
cover any incidental or consequential damages, commercial loss or any other 
damages from the use of this product. 

The warranty is invalidated by any non-factory modification which will immediately 
terminate all liabilities on Woodley Equipment Company Ltd for the products or 
damages cause by its use. The buyer and its customer shall be responsible for the 
product or use of products as well as any supervision required for safety. If 
requested, the products must be returned well packed and all shipping charges must 
be paid. 

Summary of Contents for Clinispin HCT II

Page 1: ...1 User Manual...

Page 2: ...ssories 3 5 Technical Specifications 4 6 Safety Precautions 4 7 Installation 5 8 Standard Parts Listing 7 9 User Interface and Display 7 10 Rotor Installation 8 11 Operating the Centrifuge 9 12 Mainte...

Page 3: ...of micro blood NOTE Before using the centrifuge please read this user manual carefully This user manual is intended to assistwith theoperation andcareoftheunitonlyandnotitsrepair For repairplease cont...

Page 4: ...pressing STOP Check the rotor or rotor lid and fasten it if required The rotors must be loaded symmetrically Each tube should be counter balanced by another tube of the same weight The rotor lid must...

Page 5: ...efore using cleaning or decontamination methods other than those stated in this manual please contact Woodley Equipment Company to ensure that the intended method will not damage the centrifuge The po...

Page 6: ...hile connecting the power adaptor 3 The sticker on the body and jack indicate the position of the rib on the jack While fitting the adaptor to the unit please match the sticker position as shown in th...

Page 7: ...e ongoing operation Lid opens automatically after rotor comes to a stop 3 SPEED TIME BUTTON Long press button to select TIME mode Press or to increase or decrease the values 4 Increase speed and time...

Page 8: ...n the rotor 4 After properly fastening the rotor place the rotor lid on the rotor lid nut by hand and rotate the rotor lid nut clockwise NOTE 1 Check the rotor is rmly tightened before running the nex...

Page 9: ...or 5 minutes To adjust this after closing the centrifuge lid press the SPEED button to select the speed setting Press or to increase or decrease values in intervals of 100 RPM The minimum and maximum...

Page 10: ...fugation Err 55 will be shown on the display First correct the imbalance load using the method described in Section 10 of this manual After correcting the imbalance switch the centrifuge off and on Th...

Page 11: ...or codes below Error Problem Solution No display No mains power connection Check power and properly plug in the power adaptor at both ends Power failure Check the main fuse Improper connection Connect...

Page 12: ...ration 14 Warranty This product is warranted to be free from defects in material and workmanship for a period of two 2 years from date of purchase Your product will be duly repaired upon prompt notifi...

Page 13: ...rned for service should be returned in the original packaging or suitable carton padded to avoid damage Woodley Equipment Company is not responsible for any damage incurred by improper packaging This...

Page 14: ...14 Old Station Park Buildings St John Street Horwich Bolton BL6 7NY UK Tel 44 0 1204 669033 Email sales woodleyequipment com Web www woodleyequipment com...

Reviews: