length, follow the procedure on the same page to change the 
preset time.

 

Cycle parameters cannot be changed

 

o

 

If cycle selection is locked on one cycle, press the UNLOCK 
button for 5 seconds. Then, press the GEAR button and follow 
the instructions on page 6 of this manual.

 

o

 

If different saved cycles can be selected but not modified, the 
centrifuge is in Preset Lock mode. Press the UNLOCK button for 
5 seconds until two beeps are heard, then again until the next 
two beeps. You should now be able to change cycle parameters.

 

The centrifuge does not unlock after a 
run is completed

 

o

 

Wait until the rotor has come to a complete stop. If the lid knob 
still cannot be rotated, press the UNLOCK button and try again.

 

o

 

If no LED light is on, the 

unit is not powered and the lid will 
not unlock by conventional means. 
Remove the latch label and use a 
pen to manually disengage the 
locking mechanism. Pull the 
mechanism towards the control 
panel and then unlatch and open 
the lid. 

o

 

If the unit is damaged, contact Technical Support for assistance.

 

The lid does not open

 

o

 

Ensure that the lid knob is turned fully counterclockwise.

 

o

 

If the knob cannot be turned counterclockwise, turn it 
fully clockwise, press UNLOCK, and turn counterclockwise.

 

o

 

If the lid remains locked after this and will not unlock, the 
electronics may have been damaged. Contact Technical 
Support for assistance.

 

Clicking noise during braking gets loud

 

Make sure that the screw in the centre of the rotor is tight.

 

Lid does not stay up

 

Tighten the centre screw on the lid hinge.

 

 

Summary of Contents for Clinispin HORIZON 12 Flex

Page 1: ...Clinispin HORIZON 12 Flex 24 Flex Operator s Manual...

Page 2: ......

Page 3: ...e without Saving 4 Cycle Lock 5 Single Cycle Lock 5 Preset Lock 5 Setting or Modifying a Saved Cycle 6 Deleting a Cycle 6 Reset to Factory Settings 7 Settings 7 Operation 7 Balancing Loads 8 Care and...

Page 4: ...dible alert indicates the completion of the cycle The audible alert can bemuted o Cool Flow design prevents overheating of samples by using ambient air to keep specimens at room temperature o The tube...

Page 5: ...energy and if not installed and used in accordance with this operator manual may cause interference to radio communications Operation of this equipment in a residential area may cause interference in...

Page 6: ...or Chemistry tubes 2 Coag PPP This setting is factory preset for Coagulation or Platelet Poor Plasma PPP 3 Urine This setting is factory preset for urine tubes Start Begins running the cycle displayed...

Page 7: ...op screen displays the speed Setting by G force Press and hold the RCF xg button while changing the displayed setting on the top screen using the up and down buttons next to it The RPM will automatica...

Page 8: ...and a flash of the lid light will confirm that cycle selection is locked Cancel Single Cycle Lock To re enable cycle selection with lid open press and hold the UNLOCK button for 5 seconds Two beeps an...

Page 9: ...ons next to the bottom TIME display to adjust time Change Brake Values Navigate with the down key next to the top screen until it shows Brake To change the value shown on the BRAKE display use the up...

Page 10: ...ATION o Place the tubes into the tube holders Be sure to follow the rules for balanced loads as listed in the next section o Close the lid and turn the lid knob clockwise to its complete stop position...

Page 11: ...when loading the rotor If an odd number of samples is to be spun fill a tube with water to match the weight of the unpaired sample and place it across from this sample Opposing tube holders must be e...

Page 12: ...the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there is a spillage or...

Page 13: ...kets have any damage evenslight safely dispose of them and replace them o If the rotor appears damaged contact Technical Support for further assistance Excessive noise or vibration when the centrifuge...

Page 14: ...If the lid knob still cannot be rotated press the UNLOCK button and tryagain o If no LED light is on the unit is not powered and the lid will not unlock by conventional means Remove the latch label a...

Page 15: ...59 dB A Environmental Range 16 32 degree C 16 32 degree C Voltage 95 253 VAC 95 253 VAC Frequency 50 60 Hz 50 60 Hz Power Requirement 415 Watts 415 Watts Centrifuge Motor H P Brushless H P Brushless M...

Page 16: ...their waste equipment by handing it over to a designated collection point for the recycling of waste electrical and electronic equipment The separate collection and recycling of your waste equipment a...

Reviews: