WARNING: For the safety of both the operator and service personnel, care should be taken when using 

this  centrifuge  if  handling  substances  that  are  known  to  be  toxic,  radioactive  or  contaminated  with 
pathogenic microorganisms. Use appropriate personal protection equipment (PPE). When Risk Group II 
materials are used, (as identified in the World Health Organization 

“Laboratory

 Bio-Safety 

Manual”),

 a Bio- 

Seal should be employed. In the event that materials of a higher risk group are being used, more than one 
level  of  protection  must  be  provided.  The  use  of  flammable  or  explosive  materials  as  well  as  those 
materials which have a vigorous chemical reaction is prohibited. 

 

   

Unplug the centrifuge before cleaning or performing maintenance. 

 

   

WARNING: Inspect centrifuge for cracks or physical damage to cabinet, lid, rotor or tube holders. Damage 

may result in unsafe operation. Discontinue use until repairs have been performed. 

 

   

This equipment generates, uses and can radiate radio frequency energy and if not installed and used in 

accordance with this operator manual, may cause interference to radio communications. 

 

   

Operation of this equipment in a residential area may cause interference, in which case the user will be 

required to correct the interference at their own expense. 

 
 

I

NITIAL 

S

ETUP 

o

 

Unpack and verify that all the following are included:

 

 

Centrifuge

 

 

Power cord

 

 

Tube holders

 

 

Operating manual

 

 

o

 

Setup the centrifuge on flat and level surface. A benchtop clearance height of 54cm (

21”

) is required to open 

the lid.

 

 

o

 

The centrifuge should have  15cm (

6”

) of clear space around the  centrifuge. Proper ventilation is necessary to 

prevent the overheating of samples as well as premature failure of the centrifuge. Choose an area which allows 
unencumbered air flow, and where the temperature remains between 16°C and 32°C.

 

 

o

 

No hazardous material shall be permitted in the clearance envelope during operation.

 

o

 

The operator time within the envelope shall be limited to the time necessary for loading, unloading and 
centrifuge operation only.

 

 

o

 

Plug the line cord into the centrifuge.

 

o

 

Plug the line cord into an approved electrical outlet.

 

 

 

   

B

E  SURE  THE  ELECTRICAL  OUTLET  IS  ALWAYS  ACCESSIBLE  AS  THE  LINE  CORD  IS THE  MEANS  OF  EMERGENCY

 

D

ISCONNECTION

!

 

Summary of Contents for Clinispin HORIZON 12 Flex

Page 1: ...Clinispin HORIZON 12 Flex 24 Flex Operator s Manual...

Page 2: ......

Page 3: ...e without Saving 4 Cycle Lock 5 Single Cycle Lock 5 Preset Lock 5 Setting or Modifying a Saved Cycle 6 Deleting a Cycle 6 Reset to Factory Settings 7 Settings 7 Operation 7 Balancing Loads 8 Care and...

Page 4: ...dible alert indicates the completion of the cycle The audible alert can bemuted o Cool Flow design prevents overheating of samples by using ambient air to keep specimens at room temperature o The tube...

Page 5: ...energy and if not installed and used in accordance with this operator manual may cause interference to radio communications Operation of this equipment in a residential area may cause interference in...

Page 6: ...or Chemistry tubes 2 Coag PPP This setting is factory preset for Coagulation or Platelet Poor Plasma PPP 3 Urine This setting is factory preset for urine tubes Start Begins running the cycle displayed...

Page 7: ...op screen displays the speed Setting by G force Press and hold the RCF xg button while changing the displayed setting on the top screen using the up and down buttons next to it The RPM will automatica...

Page 8: ...and a flash of the lid light will confirm that cycle selection is locked Cancel Single Cycle Lock To re enable cycle selection with lid open press and hold the UNLOCK button for 5 seconds Two beeps an...

Page 9: ...ons next to the bottom TIME display to adjust time Change Brake Values Navigate with the down key next to the top screen until it shows Brake To change the value shown on the BRAKE display use the up...

Page 10: ...ATION o Place the tubes into the tube holders Be sure to follow the rules for balanced loads as listed in the next section o Close the lid and turn the lid knob clockwise to its complete stop position...

Page 11: ...when loading the rotor If an odd number of samples is to be spun fill a tube with water to match the weight of the unpaired sample and place it across from this sample Opposing tube holders must be e...

Page 12: ...the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there is a spillage or...

Page 13: ...kets have any damage evenslight safely dispose of them and replace them o If the rotor appears damaged contact Technical Support for further assistance Excessive noise or vibration when the centrifuge...

Page 14: ...If the lid knob still cannot be rotated press the UNLOCK button and tryagain o If no LED light is on the unit is not powered and the lid will not unlock by conventional means Remove the latch label a...

Page 15: ...59 dB A Environmental Range 16 32 degree C 16 32 degree C Voltage 95 253 VAC 95 253 VAC Frequency 50 60 Hz 50 60 Hz Power Requirement 415 Watts 415 Watts Centrifuge Motor H P Brushless H P Brushless M...