background image

Operator’s Manual

CLINISPIN MPC M

ulti 

P

urpose 

C

entrifuge

Rev. 1.0

Summary of Contents for Clinispin MPC

Page 1: ...Operator s Manual CLINISPIN MPC Multi Purpose Centrifuge Rev 1 0...

Page 2: ......

Page 3: ...up Location Initial Setup Procedure Operation Rotor Configuration Filling of Capillary Tubes Rotor Installation Control Panel Functions Advanced User Settings Care and Maintenance Troubleshooting page...

Page 4: ...MODEL DESCRIPTION The Clinispin Multi Purpose Centrifuge MPC is a high speed centrifuge equipped with a maintenance free drive digital display and simple interface for silent efficient operation It i...

Page 5: ...bient Temperature during operation 4 C to 40 C Maximum Relative Air Humidity 80 Noise Level 60 dB Electrical Requirements Input Voltage 110 240 VDC 50 60 Hz Output Voltage 24VDC 6 25 A Power Consumpti...

Page 6: ...d Button RPM indicator LED Increment button Serum Plasma indicator LED Serum Plasma button preset HCT QBC button preset Urine button preset HCT QBC indicator LED Speed display Time Display RCF indicat...

Page 7: ...rs and spare parts Centrifuge may be used for the specified applications only It must not be operated in a hazardous or flammable environment and must not be used to centrifuge explosive or highly rea...

Page 8: ...s material shall be permitted in the clearance envelope during operation The operator time within the envelope shall be limited to the time necessary for loading uploading and centrifuge operation onl...

Page 9: ...trifuge A smooth whirring sound should be heard If there are any loud or unusual sounds stop the centrifuge by pressing the OPEN STOP button immediately and refer to the section on troubleshooting pag...

Page 10: ...d IF A PROBLEM IS FOUND DURING A SPIN THAT REQUIRES THE CENTRIFUGE TO BE SHUT DOWN PRESS THE OPEN STOP BUTTON IMMEDIATELY 8 Once the set speed is reached the timer will begin the remaining time will b...

Page 11: ...proper tube loading 1 Opposing tube carriers must be identical 2 Opposing tube carriers must be empty or loaded with an equal number of equally weighted samples 3 If an odd number of samples is to be...

Page 12: ...nd of the capillary Insert the capillary tube vertically into the seal up to the point where the end of the capillary touches the bottom of the seal plate Slightly tilt the capillary tube to the side...

Page 13: ...d load vertically on the motor shaft 2 Place the rotor nut in the center hole of the rotor onto the motor shaft 3 Put T Allen Key in rotor nut turn clockwise to tighten and anti clockwise to loosen ro...

Page 14: ...HCT QBC For HCT rotor This HCT QBC button has the pre defined speed time value of 12000 RPM 5 minutes respectively Single press the HCT QBC button to start operation at 12000 RPM for 5 minutes of tim...

Page 15: ...will run for a maximum of 99 minutes and will then shut off automatically The ongoing centrifugation may also be terminated by pressing the STOP OPEN button The remaining time after the termination o...

Page 16: ...the centrifuge lid opens release the emergency release cap back in the hole Emergency lid release pin Pull Down using any tool Lid unlock Imbalance Detection This centrifuge is equipped with imbalance...

Page 17: ...here are no brushes to replace and it should not need routine servicing for the life of the centrifuge The electrical components are selected for high reliability and should not need service 3 Keep th...

Page 18: ...epresentative Solution Rotor not loaded symmetrically imbalance detected Restart centrifuge and load rotor symmetrically Solution Rotor jammed remove any obstruction to allow free rotor movement Motor...

Page 19: ......

Page 20: ...acement parts or accessories Old Station Park Buildings St John s Street Horwich Bolton BL6 7NY United Kingdom Tel 44 0 1204 669033 Fax 44 0 1204 669034 www woodleyequipment com Email sales woodleyequ...

Reviews: