page 10
ROTOR CONFIGURATION
This rotor is designed to hold up to 24 capillary tubes,
up to size of OD: 1.5mm & Length: 75mm
Follow the procedure below for proper tube loading.
1. Opposing tube carriers must be identical.
2. Opposing tube carriers must be empty or loaded with an equal number of
equally weighted samples.
3. If an odd number of samples is to be spun, use a water-filled tube to mate
with the unpaired one.
This rotor is designed to hold up to 12 test tubes,
up to size of volume of 2ml
Follow the procedure below for proper tube loading.
1. Opposing tube carriers must be identical.
2. Opposing tube carriers must be empty or loaded with an equal number of
equally weighted samples.
3. If an odd number of samples is to be spun, use a water-filled tube to mate
with the unpaired one.
Twenty four (24) Carrier Horizontal Rotor (pre-installed):
Twelve (12) Carrier Fixed Angle Rotor (additional):
Your centrifuge must contain a balanced load in order to
work properly. To ensure that the load is balanced, keep
these below rules in mind when inserting capillary tubes
in to the rotor.
Your centrifuge must contain a balanced load in order to
work properly. To ensure that the load is balanced, keep
these below rules in mind when inserting tubes
in to the rotor.