page 13
CONTROL PANEL FUNCTIONS:
Predefined Pre-sets:
The centrifuge has Predefined Pre-sets for easy accessing of commonly used
operations. After closing the centrifuge lid, a single press of any pre-set
button will display the pre-set ‘speed’ and ‘time’ to be used. Below are the 3
available pre-sets buttons.
1) SERUM / PLASMA (For Microtube rotor)
This “SERUM / PLASMA” button has the pre-defined speed & time value of 12000
RPM & 2 minutes respectively. Single press the “SERUM / PLASMA” button to start
operation at 12000 RPM for 2 minutes of time. Once the button is pressed, the
predefined speed and time will appear in the display.
2) HCT / QBC (For HCT rotor)
This “HCT / QBC” button has the pre-defined speed & time value of 12000 RPM &
5 minutes respectively. Single press the “HCT / QBC” button to start operation at
12000 RPM for 5 minutes of time. Once the button is pressed, the predefined speed
and time will appear in the speed and time buttons display. This may be amended
This may be amended manually
if required with the ‘time’ and ‘speed’ buttons.
manually if required with the ‘time’ and ‘speed’ buttons.
3) URINE (For Microtube rotor)
This “URINE” button has the pre-defined speed & time value of 1000 RPM &
1 minute respectively. Single press the
to start operation at 1000
“URINE” button
RPM for 1 minute of time. Once the button is pressed, the predefined speed and time
will appear in the display. This may be amended manually if required with the ‘time’
and ‘speed’ buttons.