37
Section 7 Error Codes and Troubleshooting
7.1
Error Code 0101
Multi-way Switch Failure: Mainly due to component damage on the acquisition circuit.
This error is very rare and if this problem occurs, please contact Woodley Equipment
Company.
7.2
Error Code 0102
+12V Power Supply Abnormal: Usually due to damage of circuit components or abnormal
output of power supply board. This error is very rare and if this problem occurs, please
contact Woodley Equipment Company.
7.3
Error Code 0103
-12V Power Supply Abnormal: Usually due to damage of circuit components or abnormal
output of power supply board. This error is very rare and if this problem occurs, please
contact Woodley Equipment Company.
7.4
Error Code 0105
AD (AD Fault) Transformation Fault: Usually due to AD conversion chip damage, no
conversion signal output. This error is very rare and if this problem occurs, please contact
Woodley Equipment Company.
7.5
Error Code 0107
Optical Module Fault:
7.5.1
Reasons
1)
Due to the dirty surface of the optical lens, the light intensity attenuation is large.
2)
The problem of optical signal connection line leads to the attenuation of the light intensity
collected.
3)
The problem of the circuit board card of the collection signal appears and the collection
signal becomes weak.
4)
Xenon lamp is not exposed.
5)
The filter is mouldy or when the filter adhesive layer is corroded by alcohol, which leads
to lower transmittance of the filter after a long time.