41
4)
Replace the FPGA cable
1. Wipe the FPC cable with anhydrous
ethanol and take off the hot melt glue on
the upper and lower connector joint.
Remove the FPGA cable after opening
the clasp of connector. Use anhydrous
ethanol to wipe the conductor pins of the
cable, then plug back into the analyser
and restart the machine.
5)
Replace xenon lamp exposure control line.
1. Confirm the model of the xenon lamp,
change to a new xenon lamp connect line
if there is a black connector as in the
figure on the right. Then, assemble the
analyser again.
Check if the analyser recovers to flash on
the position of the optical lens.