43
1. Take out the connector of xenon lamp
connect line on the control board.
Use the debugging function to make an
exposure test.
If there is a problem with the lamp,
proceed with the following steps.
2. Remove two screws from the bottom of
the front panel. As shown in the figure.
3. Remove two screws at the front of the
front panel. As shown in the figure.
Remove the front panel.
4. Remove the xenon lamp and the
positioning wire harness connector
after cutting the ribbons with the pliers,
as shown in the figure.