7.10
Error Code 0208, 0209, 0210 & 0211
Error 0208 Insufficient Sample: The absorbance difference between the two corresponding
compared cuvettes is less than threshold.
Error 0209 Insufficient Dilution: The absorbance difference between the two
corresponding compared cuvettes is more than threshold.
Error 0210 Mixed Solution Not in Place: The test results of specific analytes are less than
the threshold.
Error 0211 Mixed Unevenness: The absorbance difference between the corresponding two
compared cuvettes is more than threshold.
Collect the images of the error disc under the standard position and contact Woodley
Equipment Company.
7.10.1 Reasons
Error 0208:
1)
Insufficient sample volume.
2)
The sample does not flow through the sample channel (clotted sample, etc.)
3)
At the sample adding port, the sample is thrown out, resulting in insufficient sample volume.
4)
The liquid flows into the other chambers and doesn't flow into the waste hole.
Error 0209:
1)
Insufficient diluent has been added.
2)
The diluent is in the quantification stage, the remaining diluent leaks, so no diluent flows
into the compared cuvette.
Error 0210:
1) The blood sample has severe lipemia and cannot flow through the flow channel and reach the
correct reagent cuvette.
Error 0211:
1)
Haemolysis samples have a darker colour and higher absorbance of the blood, so it is very
easy to appear that the absorbance deviation of the compared cuvette and the reagent cuvette
exceeds the threshold.