54
2)
Lipid is heterogeneous, the absorbance difference between the compared cuvette and the
reagent cuvette is easier to exceed the threshold.
3)
Icterus is very similar to haemolysis but it is not easy to appear 0211 unless it is very serious
4)
Samples that cannot be tested by 0210 will get error 0211. When blood flows through the
sample flow channel, it doesn't flow through directly at one time, resulting in the error code.
5)
In addition, the slight 0210 will appear that a part of the mixed liquid flows into the reagent
hole first and then the other blood sample flows again a short time later. It means that there
is a sample that does not pass through the channel in the high-speed centrifugal stage at the
front. The subsequent centrifugation process starts again so there is an interval between the
time of the different reagent holes. If the deviation of the absorbance of the compared
cuvette and the reagent cuvette is greater than the threshold but the result of setting item
also under the requirements, no error will be reported.
7.10.2
Troubleshooting
Please follow the steps in order. After each step is finished, retest whether the fault is cleared
or not. If not, proceed to the next step.
1)
Use a standard serum test. Stopping error reporting is a sample problem, otherwise please
confirm uploading data and contact Woodley Equipment Company.
7.11
Error Code 0212
Disc Expired
7.12
Error Code 0213
Temperature Fault:
1)
The temperature does not rise to 36°C when it finishes mixing .
2)
The temperature exceeds 37 ± 1
℃
during testing.
7.12.1
Reasons
1)
The temperature of the environment when the analyser is running does not meet the
requirement of 10-30°C .
2)
The cooling fan behind the analyser needs to be cleaned.
3)
Temperature inaccuracy due to the temperature sensor connection problem.
7.12.2
Troubleshooting
Please follow the steps in order. After each step is finished, retest whether the fault is
cleared or not. If not, proceed to the next step.
Summary of Contents for InSight V-CHEM
Page 1: ...Veterinary Chemistry Analyser Service Manual...
Page 9: ...3 Back View Right View...
Page 42: ...36...